OP900 Auto Accumulate
6:12
21 күн бұрын
OP924 New Style Calibration
2:18
2 ай бұрын
OP909 Plus Count Feature
1:13
5 ай бұрын
OP900 Dual Port Setup
1:50
8 ай бұрын
Changing Units on OPX-SS
1:18
9 ай бұрын
OP903 Enabling Print New Style
0:52
OP903 Calibration New Style
3:40
10 ай бұрын
Updating New Style OP903
1:31
11 ай бұрын
OP901 Fixing NNNN on Compression
0:35
T903 6 Pad calibration
5:45
Жыл бұрын
OP900B Setting Decimals in Grams
1:31
Connecting Scale to Ethernet
10:53
New TCB880 Calibration
3:30
Жыл бұрын
Пікірлер
@nixic_
@nixic_ 10 күн бұрын
Super helpful - my old ass Losco had some loose wires. Thanks for the tip!
@murugudumanojkumar5833
@murugudumanojkumar5833 2 ай бұрын
What is the use of junction box and how potentiometer in junction box will be used?
@optimascale6851
@optimascale6851 Ай бұрын
Protection: Safeguards electrical connections from environmental and mechanical damage. Organization: Provides a central point for managing electrical wiring connections, making maintenance easier. Safety: Reduces the risk of short circuits, fire hazards, or electrical shocks by containing and insulating connections. Signal Calibration: In sensor systems, potentiometers are used to fine-tune signal levels for calibration, ensuring the sensors operate within desired ranges. Adjusting Output: It may regulate output voltage or current in circuits, allowing customization of electrical parameters to match system requirements. Balancing Circuits: Potentiometers can balance circuits or adjust offsets in control systems, ensuring stable and accurate operation. Thanks, Anthony
@sharifhossain8830
@sharifhossain8830 2 ай бұрын
have formula mv to temperature?
@LuisCarlosOsuna-u8t
@LuisCarlosOsuna-u8t 2 ай бұрын
Hola buenos días disculpe para comentar por qué al prender el aparato sigue la secuencia marca error6 y se apaga
@optimascale6851
@optimascale6851 2 ай бұрын
Tal vez necesites conectarlo porque la batería está descargada. Intenta calibrarlo nuevamente. Si no funciona, es posible que tengas que abrir la parte trasera y desconectar los cables de la batería.
@jrmlp1291
@jrmlp1291 2 ай бұрын
Error 2?
@optimascale6851
@optimascale6851 2 ай бұрын
If the indicator is 10+ years old then you would need to use 10% of capacity to calibrate. If newer than that check the wiring in the platform. There is video on channel on how to fix floor scale. Follow cable to junction box in platform and open that up. Check wiring in there. Might be a wire loose or in wrong order. They should all be in the same order. Thanks, Anthony
@coleshambaugh7738
@coleshambaugh7738 3 ай бұрын
i did this and got it to say 0lbs but now its not reading any weight. It just stays at zero
@optimascale6851
@optimascale6851 3 ай бұрын
Did you make sure you put weight during the adload on the screen. If you did correctly try 1 more time. If that doesn't work you might have a problem with your OP902. You would need to replace the indicator with a new indicator like our OP909. They are $165 + tax + shipping. You can email [email protected] to purchase. Typically these OP902 have problems like this after a couple years. Thanks, Anthony
@chocoso100
@chocoso100 4 ай бұрын
Mine is broke, where i can find other one?
@optimascale6851
@optimascale6851 3 ай бұрын
email: [email protected] Send a photo of which model you have.
@AdrianMatheusVDS
@AdrianMatheusVDS 5 ай бұрын
It connects but doesn't give me any data. Does any one know why? cause` the prompt apperas but there`s no data in there
@optimascale6851
@optimascale6851 4 ай бұрын
You might need to check the indicator settings to ensure it is set to continuous and that the correct baud rate is selected. If the settings are correct, you may need a null modem cable, as pins 2 and 3 are swapped. However, if you are already using a null modem cable, try switching to a regular (non-null modem) cable. Alternatively, you can purchase a DB9 null modem adapter (male to female). The female end plugs into the DB9 male connector of the cable, which then connects to the indicator. This will also swap pins 2 and 3. Pin 2 is TX (Transmit), and Pin 3 is RX (Receive) on the DB9 connector. These items are readily available on Amazon. If none of this works, there might be an issue with the internal board. Finding the correct cables and null modem adapters can be tricky. Thanks, Anthony
@AdrianMatheusVDS
@AdrianMatheusVDS 4 ай бұрын
@@optimascale6851 I'll try, thank you dude
@alexandercosigua9699
@alexandercosigua9699 5 ай бұрын
Buen día Como configurar Para agregarle o quitarle decimales. Tengo un MODEL OP-900A MS. Gracias
@optimascale6851
@optimascale6851 5 ай бұрын
Puedes usar la configuración de CO2 para establecer la cantidad de decimales que desees.
@ngtechinfo4281
@ngtechinfo4281 7 ай бұрын
Good
@rumahkuburmarimasuk7481
@rumahkuburmarimasuk7481 7 ай бұрын
Sir I have dis model Problem cod erro6 what problem dis
@optimascale6851
@optimascale6851 7 ай бұрын
Typically try to calibrate like in the video. If that doesn't help maybe problem with main PCB. You might need to check the wiring. Here is another video on the channel about this. kzbin.info/www/bejne/qofFhWh9q7J8Y9Esi=m8ziorYWR2xXBqfO Thanks, Anthony
@jonvogel7644
@jonvogel7644 7 ай бұрын
I spent 2 hours trying to figure it out on my own with the manual and in less than 2 minutes of video... I've finally got it working! Thank you!
@optimascale6851
@optimascale6851 7 ай бұрын
Glad it helped! -Anthony
@muhammadrozan1802
@muhammadrozan1802 9 ай бұрын
Thank you for the tutorial
@myaqoob6842
@myaqoob6842 9 ай бұрын
Nice 👍
@falalumahmood
@falalumahmood 9 ай бұрын
Please can u install floor scale without junction box
@optimascale6851
@optimascale6851 9 ай бұрын
No since there is 4 load cells. The indicators only have room for 1 connection. You can connect one load cells directly to indicator but that is all. Other option would be to buy OP903 which has 4 5 pin connectors and attach a 5 pin connector to each load cell and connect that to each port of 903. Thanks, Anthony
@lolofranco5975
@lolofranco5975 10 ай бұрын
Hola estimado amigo,podrías indicarme las configuraciones para poder calibrar un indicador de peso para ganado op-900 ?
@optimascale6851
@optimascale6851 10 ай бұрын
C01-2 C02-0 C03-1 C04-5000 C05-1 C06-1 Tengo un manual en español en nuestro sitio web aquí. www.optimascale.com/manuals Thanks, Anthony
@lolofranco5975
@lolofranco5975 10 ай бұрын
@@optimascale6851 hermano,con mucha sinceridad agradezco el tiempo que te has tomado para contestarme.Te comento rápido,que yo calibré esa báscula la cual se usa para pesar ganado y el punto está que una vez calibrada cuando se realiza prueba de peso con una res el peso se mantiene oscilante,no estable.podría ser falta de configuración en la sensibilidad?
@ejray2k
@ejray2k 11 ай бұрын
This was very helpful. Thanks!
@Bartes_terazi_sistemleri
@Bartes_terazi_sistemleri 11 ай бұрын
Hello, where can you buy the white color printer and program?
@optimascale6851
@optimascale6851 10 ай бұрын
Sorry for taking long time to respond. Email [email protected] to place your order. Anthony
@krizianannferranco3022
@krizianannferranco3022 Жыл бұрын
Hm
@thabisovuyohlalukana5037
@thabisovuyohlalukana5037 Жыл бұрын
I've tried this but it just doesn't work for me. It connects but doesn't give me any data.
@jerven1128
@jerven1128 9 ай бұрын
have you checked your port?
@morkovija
@morkovija Жыл бұрын
thank you for detailed explanation. Surprisingly there arent that many videos on the subject
@hotjulysun
@hotjulysun Жыл бұрын
Ours won’t zero out.
@optimascale6851
@optimascale6851 Жыл бұрын
If your readout is older than 2-3 years old there might be a problem with the readout. You might need to replace. We discontinued the OP902 and replaced with OP909JR. Ask for the 4 to 5 pin adapter. email: [email protected] to place order. Otherwise it could be damaged cable someone is sales can help you with that. Anthony
@TheIntertainers
@TheIntertainers Жыл бұрын
What position should the lever on the handle be when you install it?
@optimascale6851
@optimascale6851 Жыл бұрын
About 45 degrees
@TheIntertainers
@TheIntertainers Жыл бұрын
@@optimascale6851 I'm sorry, I meant the three-position operational handle lever
@wadesherman1477
@wadesherman1477 Жыл бұрын
Legend for making this!! Thanks
@magnusawukukobina8295
@magnusawukukobina8295 Жыл бұрын
how can get that indicator...i need 1
@optimascale6851
@optimascale6851 Жыл бұрын
Email [email protected] to purchase.
@OneArmCarGuy
@OneArmCarGuy Жыл бұрын
Thank you
@ShivaKumar-zp3xu
@ShivaKumar-zp3xu Жыл бұрын
Now they are good values so load cells are good How we knew if they are bad if ot comes morr resistamce??? 😊
@optimascale6851
@optimascale6851 Жыл бұрын
Only way to check is add voltage over red and black and measure output resistance as weight is being added.
@jyvben1520
@jyvben1520 Жыл бұрын
for windows, the program used was Putty, this also works on linux, where the port(s) are /dev/tty.... e.g. ttyACM0, ttyUSB0 the linux user (not root) might need to be part of the dialout group, do the following in a terminal (need sudo rights) sudo adduser $USER dialout (where $USER points to the current logged in active user, or replace it with your logon name )
@GrandMughal-369.
@GrandMughal-369. Жыл бұрын
Kindly make single ponit loadcell making compelete process vidio.
@i_like_lasers
@i_like_lasers Жыл бұрын
amazing. big thanks
@bluebox9971
@bluebox9971 Жыл бұрын
question, there is a wireless version of this, what is the name of the board used for making it wireless?
@optimascale6851
@optimascale6851 Жыл бұрын
You can use what you want. We have used the digi xbee in the past.
@optimascale6851
@optimascale6851 Жыл бұрын
Anthony
@AaronMadison-t5q
@AaronMadison-t5q Жыл бұрын
Where can I purchase a replacement battery for my pallet jack, I can't seem to find them anywhere.
@optimascale6851
@optimascale6851 Жыл бұрын
email [email protected] with which color battery you have we have in stock.
@محمد-ظ3ذ5ص
@محمد-ظ3ذ5ص Жыл бұрын
How can 12 cells be connected together?
@optimascale6851
@optimascale6851 Жыл бұрын
You can use a 12 port junction box. We only have 10 ports here. Second option is to use to SECTIONAL junction boxes for example 2 8 port junction boxes where 1 goes into into to the input of the other junction box. Diagram would be readout -> junction box 1 (4 load cells) -> junction box 2 (8 load cells).
@rondemus9736
@rondemus9736 Жыл бұрын
Necesito el esquema de como soldarlo
@optimascale6851
@optimascale6851 Жыл бұрын
Rojo-TX Verde - RX Negro-tierra
@ethanengle7927
@ethanengle7927 Жыл бұрын
Hey man, great videos. I do have a question about error 3 and 4 for the 900b ex. As well as when a scale head is powered on and just has dashes and won’t let me do anything.
@optimascale6851
@optimascale6851 Жыл бұрын
error 3 is input signal is negative. This usually happens when the green and white wires are swap and giving a negative signal. You can check the C15 setting to see how it responds to weight. The video on the channel called how to fix a floor scale explains more about this. error 4 happens when weight is unstable. Usually customers put too many decimals that their scale can't handle. Try to use less decimals and try out. If it still happens then probably damaged cable to readout. Check C15 setting as well. I haven't seen the dashes probably overload or short in cable. Try to go into settings and recalibrate. email: [email protected] for any specific questions you have.
@stuartmorales308
@stuartmorales308 Жыл бұрын
How do you turn on the Battery Indicator??
@optimascale6851
@optimascale6851 Жыл бұрын
It should turn on when pressing the power button up front. Make sure power cables connected. Anthony
@kamalabdalla7326
@kamalabdalla7326 2 жыл бұрын
Nice
@OrVardi-ik9pd
@OrVardi-ik9pd 2 жыл бұрын
is there a way to have this print in Excel
@optimascale6851
@optimascale6851 2 жыл бұрын
You would need some kind of software. Here is a video below on an example software we make. kzbin.info/www/bejne/g5uvXmZmjKukg68
@SaurabhPatel-qx9sg
@SaurabhPatel-qx9sg 2 жыл бұрын
Thank you for your demo for calibration. It worked for the ULINE floor scale too.
@optimascale6851
@optimascale6851 2 жыл бұрын
Yes, we ship for ULINE as well. Sometimes they rebadge our scales as their own or find another supplier and do the same. All scales work the same so the any videos on this channel will help. Anthony
@timothyabanigbe7913
@timothyabanigbe7913 2 жыл бұрын
Thank you
@hoangthe8196
@hoangthe8196 2 жыл бұрын
where download Software ?
@optimascale6851
@optimascale6851 2 жыл бұрын
We sell this software. You can email: [email protected] to purchase. Mentions you want to buy data logging software. They can give you a price.
@hoangthe8196
@hoangthe8196 2 жыл бұрын
@@optimascale6851 thank you .
@Mkhayra
@Mkhayra 2 жыл бұрын
I like
@daleskiballadares8387
@daleskiballadares8387 2 жыл бұрын
Espani
@Ziggs318
@Ziggs318 2 жыл бұрын
Love your videos bro. I work for Data Weighing Systems out in Wood Dale IL and your videos help me a lot. Question, are those batteries rechargeable or will they need to be replaced every year or so?
@optimascale6851
@optimascale6851 2 жыл бұрын
These are reachable. Same as a phone they are not meant to be replaced. These batteries should last from 3-5 years depending on how often they are used. Anthony
@MohamedAslanAffan
@MohamedAslanAffan 2 жыл бұрын
Very useful 👌.. i want to know what happens when we calibrate a weighing scale.changes occurs in indicator? or in load cell?
@optimascale6851
@optimascale6851 2 жыл бұрын
Calibration is stored on the indicator. Load cell is just a sensor. If you take the calibrated indicator and switch load cells to a different load cell of same capacity it will weigh the same.
@annonymousme2528
@annonymousme2528 2 жыл бұрын
I was wondering if you can help with my scales , my scales were send out and working from who sent them but upon receiving them they didn't work, I pulled them apart to find what looks like a wire soldered in wrong spot could this be why they say unstable
@annonymousme2528
@annonymousme2528 2 жыл бұрын
I turn on my scale and all is says is "un-st then after a moment fluctuates in weight to 14kg and then goes to o-ld " it doesn't start up normally it just says unst ...
@optimascale6851
@optimascale6851 2 жыл бұрын
You can watch my video on how to fix a floor scale for a more detailed answered. kzbin.info/www/bejne/rZS9iHZ8ZbeBZ7c It could be wire in the wrong location, maybe needs to be recalibrated or bad load cell. Only way to tell is making sure all wiring is correct and tight and recalibrate. If there is water in the junction box then that will cause fluctuating weight. You will need to replace that junction box. You can email [email protected] for junction box replacement or if you find a bad load cell. Thanks, Anthony
@larryaltavilla1472
@larryaltavilla1472 2 жыл бұрын
WE HAVE STRUGGLED WITH THIS INDICATOR FOR YEARS.
@optimascale6851
@optimascale6851 2 жыл бұрын
Yes it is difficult. If you have a problem with one port you can check the F4 raw data for that connector and see what is going on. You can email: [email protected] with your problem for more help.
@fakboii6988
@fakboii6988 2 жыл бұрын
Error 4? Plis
@fakboii6988
@fakboii6988 2 жыл бұрын
T scale err4
@optimascale6851
@optimascale6851 2 жыл бұрын
Happens when weight is not stable. Make sure your test weight isn't moving. If it is not moving then check the wiring in the junction box in the platform. It could be 1 wire loose. Watch my how to fix a floor scale video on more details. Thanks, Anthony
@fakboii6988
@fakboii6988 2 жыл бұрын
@@optimascale6851 thank you, for showing me how to overcome it
@linloadcell158
@linloadcell158 2 жыл бұрын
700Ω
@optimascale6851
@optimascale6851 2 жыл бұрын
Check the green and white wires and check that ohm reading. Then Green and black should be around the same. Load cells usually don't go bad. Try to rewire and check the C15 setting. Check out how to fix a floor scale video for more details. Thanks, Anthony
@linloadcell158
@linloadcell158 2 жыл бұрын
Load cells