Good morning sir I hope you are doing very well What can explain the erroneous results when I pass the standard as samples or even the samples despite that (the aspiration / lamp control / filter control / the calibration for the recording of factors) are all normal on the Spinlab
@esaugalvez Жыл бұрын
You must use Cristal extremely clean tubes when running samples. For example if you run the same glucose sample 3 times in 3 different tubes, you must obtain minimum variation. If variation is more than 10 in each sample then you prepare in 1 tube 1,500 ul reagent plus 15ul serum and run it 3 times, in this case If variation is less than 10 then the problem are the tubes.
@ClaudiasalomeCastillopadilla Жыл бұрын
Das mantenimiento o donde te contactan ?
@esaugalvez Жыл бұрын
Si doy mantenimiento. En el perfil de la página viene mi correo. Puede escribirme ahí y explicarme que equipo es. Los servicios los hago en mi taller. Es decir, todo lo manejo por envíos, paqueterías. Saludos
@محمدالحمادي-ز5ف3 ай бұрын
لوسمحت أخي ماهو سبب هذا العطل Errorcode: 0
@esaugalvez2 ай бұрын
Send me a picture of the error. Ing.esau.galvez@gmail.com The answer will be here anyway.