Interesting video, thank you. Do you have a fix for a Visionaire 5. It beeps 4 times, indicating high pressure and the compressor belches very loudly. I think it is likely to be the sieve beds. What is the best way to service them?
@oxygenalliance10 ай бұрын
Firstly make sure the solenoid valves are functioning properly by testing their magnetism using a paper clip during the cycles. Check that the exhaust muffler is unrestricted. Also please would you share the oxygen purity level the oxygen concentrator is producing? But the ideal approach is to replace the beds with new one if they are contaminated. Alternatively you can refill them with fresh zeolite after cutting. Check out this video here: kzbin.info/www/bejne/gqa2ZGmbfs54i7M, demonstrating the process of refilling the sieve bed in a DeVilbiss 525 - applicable to Visionaire 5 beds as well.
@TambaAlieKpakima6 ай бұрын
Why cleaning the Valve?
@oxygenalliance4 ай бұрын
Sometimes, zeolite from the sieve beds can leak and accumulate in the valve airway paths and chambers. Cleaning the valve is one way to address this issue.