Good work. A little error, the pump master occlusion must be done by observing the descent of the liquid on a pipe of the same section of the tube in head pump (you observe 3/8 but have 1/2) . Furthermore, it is better for the descent to be 1 cm every 10 sec, the load loss is negligible but spallation and hemolysis are reduced.
@lobnasaleh11537 жыл бұрын
I like this stockert machine
@ellismurdock47982 жыл бұрын
Do you use hemoconcentrator on every case?
@lobnasaleh11537 жыл бұрын
A lot of recircular lines
@pritamnandy84493 жыл бұрын
What was the aortic line pressure during checking the cardioplegia pump Occlusion??
@pritamnandy84493 жыл бұрын
Please please make another video on cardioplegia pump Occlusion set
@harveypraz79623 жыл бұрын
@@pritamnandy8449 is cardioplegia pump new technology? Cause I remember a while ago cold potassium solution was injected directly into the heart by the surgeon.
@naveed42742 жыл бұрын
@@harveypraz7962 no it’s not a new technology Plegia given through pump to measure pressures and mixing fresh oxygenated blood.
@adre9812 жыл бұрын
@@harveypraz7962 No It's been a while. Cristalloid or Blood cardioplegia are infused by a pump that takes Blood directly from the oxygenator. You have to monitor the pressure of infusion depending on what kind of site of cpl infusion (antegrade retrograde or selective). Cheers ,😁