GE AMX-II, Mobile 110kV XRAY System Teardown

  Рет қаралды 5,360

Kaizer Power Electronics

Kaizer Power Electronics

Күн бұрын

Пікірлер: 13
@markhodgson2348
@markhodgson2348 10 ай бұрын
It's such a shame that it wasn't complete ,but it's very nice to see an analogue solution to a problem
@SeanBZA
@SeanBZA 6 жыл бұрын
Capacitors behind the SCR heatsink is there for commutation, using one SCR to turn off the one conducting, via the capacitor and a power resistor. The series resistors are there to limit current rise, but this is a standard 2 SCR half bridge driver, where you have a power device that handles the full load current, and another that is used to turn the conducting one off by reversing the anode cathode voltage via the capacitor transferring the current flow from the main device to the aux one. Then the aux one has a low current flowing through it, and when the main one fires again it is commutated again making it ready to switch the main one off. Done because SCR's are really robust devices, but power transistors then were fragile and blew up easy, but the SCR did not. That way they could get adjustable on time and not have both devices conducting on the bridge when switching. 100A SCR is pretty small, they went up to 1kA in almost the same package size, often used for electric forklift main motors.
@KaizerPowerElectronicsDk
@KaizerPowerElectronicsDk 6 жыл бұрын
Thank you for the details :)
@BenjaminEsposti
@BenjaminEsposti 6 жыл бұрын
Is that similar to some of the CFL/LED compatible mains dimmers?
@BenjaminEsposti
@BenjaminEsposti 6 жыл бұрын
That's a beautiful unit, it's busy in there but yet nicely tucked together.
@KaizerPowerElectronicsDk
@KaizerPowerElectronicsDk 6 жыл бұрын
Yes, so many small PCB or single components placed all around with wire harness going around to it. Not something you see in new equipment.
@jucalu93
@jucalu93 5 жыл бұрын
Hello I use an AMX system II to test our X-ray tube which is down. I was checking the equipment because the board for operator panel was damaged (burned) I replaced it but the equipment does not make exposition and also I saw smoke coming out from below of the tray of control boards. I am scary to dismount any cable because Im not expert on electrical schematics and I dont know how to start to check the failure because after the smoke 2 fuses of 8A in the front of equipment get damaged. If you have any idea on how to proceed I would be happy to know that. Regards
@johnsenchakinternetnetwork2025
@johnsenchakinternetnetwork2025 6 жыл бұрын
You power a flux capacitor with that transformer
@HboskO1982
@HboskO1982 6 жыл бұрын
Boy, you are too young for this mustache !
@KaizerPowerElectronicsDk
@KaizerPowerElectronicsDk 6 жыл бұрын
If you are from 82, we are from the same year ;)
@HboskO1982
@HboskO1982 6 жыл бұрын
@@KaizerPowerElectronicsDk yes, i am from '82. 😁
@KaizerPowerElectronicsDk
@KaizerPowerElectronicsDk 6 жыл бұрын
I could see that on your username
Teardown of a Eaton PowerWare 60kVA UPS System
17:40
Kaizer Power Electronics
Рет қаралды 10 М.
Siemens Polyphos 30, XRAY Inverter and Control Panel Teardown (Part 3 of 3)
25:23
Kaizer Power Electronics
Рет қаралды 4,6 М.
Жездуха 42-серия
29:26
Million Show
Рет қаралды 2,6 МЛН
OCCUPIED #shortssprintbrasil
0:37
Natan por Aí
Рет қаралды 131 МЛН
GE AMX 4 Plus Portable X Ray
7:32
Digital X-ray & Medical Imaging Equipment Channel
Рет қаралды 473
AMX Solutions Inc.
8:19
Cole Phail d/b/a Master Productions
Рет қаралды 4,7 М.
Another Look at an X-Ray tube:   Part 1
17:14
xraytonyb
Рет қаралды 9 М.
Contactors: how we power the big stuff
25:50
Technology Connections
Рет қаралды 1,1 МЛН
Defibrillator teardown
21:49
mikeselectricstuff
Рет қаралды 91 М.
50 kW Sedecal SHF-530 X-ray High Voltage Generator Inverter Teardown
15:25
Kaizer Power Electronics
Рет қаралды 3,3 М.
Rotating anode x ray tube / constructions and working principle
17:28
IMMORTAL UNIVERSE
Рет қаралды 22 М.
How to make high voltage transformer 400KV
17:13
American Tech
Рет қаралды 241 М.
70 kV / 65 kW XRAY System, Transformer and Inverter - Isolation, Day 12
8:20
Kaizer Power Electronics
Рет қаралды 11 М.
Siemens 135kv Monoblock Xray Head Teardown
14:51
Tested to Destruction
Рет қаралды 2,2 М.