ToolTop IR502 Battery Internal Resistance Tester Review/Teardown

  Рет қаралды 7,014

Kerry Wong

Kerry Wong

Күн бұрын

Пікірлер: 44
@originalmianos
@originalmianos 6 ай бұрын
Wow, I was just looking at one of these a few weeks ago. Thanks for the timely review.
@TradieTrev
@TradieTrev 6 ай бұрын
Great review as always mate! The 4 wire method is heaps better; I like how you always explain the theory of the instruments.
@marcfruchtman9473
@marcfruchtman9473 6 ай бұрын
Thank you for the review. I think it would have been nice for the tester to include a battery holder for the most common batteries seen today... perhaps with a few flexible holders for sizes that don't match. Not sure I can really appreciate this tester tho, without a "reference" standard to ensure that it is accurate.
@s14tam
@s14tam 6 ай бұрын
thankyou for the review
@TheDefpom
@TheDefpom 6 ай бұрын
@13:30 nice they have Rubycon caps in there.
@wtmayhew
@wtmayhew 6 ай бұрын
If the capacitors aren’t counterfeit…. The capacitors could be fakes even if the manufacturer bought them in good faith.
@simontay4851
@simontay4851 6 ай бұрын
If they were fakes, the brand would be Rulycon or Rudycon in the same font.
@simontay4851
@simontay4851 6 ай бұрын
But why are they 200V caps. No battery is 200V.
@wtmayhew
@wtmayhew 6 ай бұрын
@@TheDefpom Supposedly China is cracking down on trademark and intellectual property infringements. Sometimes I wonder how well that is enforced. In the past I have seen totally copied trademarks and the only way to tell is if the part does not perform to specification or if perhaps copy looks blurry or there is a misspelling somewhere else on the part.
@SeanBZA
@SeanBZA 6 ай бұрын
@@simontay4851 Margin, they need low leakage, so used good capacitors, as this is likely how the AC test is coupled through. Wonder how well protected it is against reverse voltage, and them giving the pinout means it is easy to make bespoke test lead sets as well.
@JeffinTD
@JeffinTD 6 ай бұрын
Interesting video, as always. If you had a chance to evaluate an ICharger S6, that might be an interesting topic.
@wtmayhew
@wtmayhew 6 ай бұрын
Profit margins have to be very thin on these instruments. You’d think it would not be worth the extra cost in labor and equipment to obliterate the part markings. I wonder if it really slows down other shady Chinese manufacturers from cloning their intellectual property? It is going to make repair of this instrument more difficult, but on the other hand it is probably intended to be disposable rather than repairable anyway.
@SeanBZA
@SeanBZA 6 ай бұрын
The parts will be delabelled before soldering, easy enough with a reel to run them through a laser to scan and erase the marking, and then stick a new top cover on, and reel them up again. Must be places that do this for money, you take them the reel, and get the erased one back.
@wtmayhew
@wtmayhew 6 ай бұрын
The people at Tooltop sure are militant about scrubbing off identification. Looks like they obliterated every semiconductor number and even sanded the frequency off of the crystal.
@simontay4851
@simontay4851 6 ай бұрын
Sanding the crystal is pointless. Its easy to measure the frequency. Even cheap multimeters have frequency measurement.
@V3T9
@V3T9 6 ай бұрын
Only reason is so no other chinaman copies the firmware and pcb design, reproduces the same product, but cheaper and destroys the original guys business that came up with it. Many such cases, sad. Still won't buy
@t1d100
@t1d100 6 ай бұрын
I did not understand, if the meter uses the Kelvin method for measuring resistors. You would think so, but I would have like to have seen that proven, one way or the other. Cheers.
@alanduncan3710
@alanduncan3710 6 ай бұрын
I see the battery meter has a "rel" button. Accuracy might improve if you rel'ed it out.
@tullgutten
@tullgutten 6 ай бұрын
Yes, every power up like clamp ampere meters in DC you should power on and wait a little, short/close the contacts together and then zero it to compensate any drift or error from leads
@pault6533
@pault6533 6 ай бұрын
Perhaps the reason the meter is giving you unexpected “resistance” readings for the 100 ohm resistor is that it is actually measuring impedance. Is that resistor wire wound inside?
@KerryWongBlog
@KerryWongBlog 6 ай бұрын
At 1kHz, I would expect some deviations but not 1.5 Ohms. I had verified the reading with an LCR meter at 1 kHz.
@pault6533
@pault6533 6 ай бұрын
I just tested some wire wound resistors between 50 and 680 ohms. The maximum error I got at 1 kHz was 0.5%. Equipment ET431 and error based on Zs/DCR. The impedance was always > resistance, as you might expect. Between using REL and switching to a Vishay rectangular precision resistor, I bet the measurement will have less than 1% error.
@LawpickingLocksmith
@LawpickingLocksmith 6 ай бұрын
This is surely useful for the off grid freak to weed out lazy batteries. Say your house has a 48V system and 12V lights, you can arrange the batteries to feed the lights from strong batteries rather than waste inequality inside the BMS and use a converter to reduce 48V down to 12V. Mustool is offering cheap solutions for many DIY folks. They could sell thie even cheaper but I guess this is a low volume product.
@johncoops6897
@johncoops6897 6 ай бұрын
This one is similar in price to existing products of this type and spec. This is a few $ cheaper and has a colour vs monochrome screen.
@wtmayhew
@wtmayhew 6 ай бұрын
Is that a security screw next to “V1” by the DE9 connector meant to discourage disassembly, or is it stripped by poor assembly?
@jax2571
@jax2571 2 ай бұрын
Accuracy cannot be judged by this review. The accuracy (inaccuracy) declared here is only the subjective opinion of the reviewer.
@KerryWongBlog
@KerryWongBlog 2 ай бұрын
You are correct. At the time of the review I didn't have another internal resistance tester to compare to. I later reviewed a FNIRSI HRM-10: kzbin.info/www/bejne/p33Ve6dmodGmgLc and did some comparison there.
@briansauk6837
@briansauk6837 6 ай бұрын
You can use most DC loads with a scope and do the same test. Set the load to pulse the current (you can even put a small draw as a baseline), and observe the voltage on a scope. Pretty easy to get decent results. For coin cells, you can use a signal generator instead of the dc load - just be careful not to set it to charge the battery (although with a milliamp or less, really not a problem)
@fly-lucky
@fly-lucky 5 ай бұрын
Couldn't you just use the ohms law method to cross validate the accuracy of the meter?
@KerryWongBlog
@KerryWongBlog 5 ай бұрын
As mentioned in the video, the resistance measured by the AC conductance method is slightly different so not an apple to apple comparison.
@fly-lucky
@fly-lucky 5 ай бұрын
@@KerryWongBlog I appreciate the fast response. I enjoy the reviews and product testing keep up the good work.
@DazzaDirect
@DazzaDirect 6 ай бұрын
@KerryWong thanks for the review looks like an interesting item i just got a YR1035+ for £20. ,, this IR502 is coming up as only £32 seems like amazing value,, could you please maybe review the Hanmatek DOS1104 ,, its £150 for a 4 channel Oscilloscope,, i appreciate and value your opinions
@paulcohen1555
@paulcohen1555 6 ай бұрын
To accurately measure the batteries with 4 wire system, you must separate the probe tips with an insulator and touch the battery at 4 points. That's not what I have seen in the video.
@KerryWongBlog
@KerryWongBlog 6 ай бұрын
I mentioned that as well. Given the battery terminal types of the batteries it was nearly impossible to do that.
@pault6533
@pault6533 6 ай бұрын
Not having another meter to compare it to? How about some hand calculations using the formulas you reviewed.
@KerryWongBlog
@KerryWongBlog 6 ай бұрын
AC conductance method will result in different results. I have another internal resistance meter coming and will do a comparison once I get my hands on that one.
@pault6533
@pault6533 6 ай бұрын
@@KerryWongBlog That will be awesome, I'll look forward to it.🥰
@_droid
@_droid 6 ай бұрын
DB9, heh. Somebody has been digging in the trash pile.
@johncoops6897
@johncoops6897 6 ай бұрын
The other well known IR meters use a 4-pin QX type aircraft connector, similar to what's used on Hakko soldering irons. Much bigger than the DB9 used here. I have one IR tester that is very slim and uses a standard USB type A plug and socket. The current is tiny so it doesn't matter about thin conductors.
@craigtucker777
@craigtucker777 6 ай бұрын
I'll rather burn 20bux than buy something what I can't fix
@jdmccorful
@jdmccorful 6 ай бұрын
Interesting video. I think the add on banggood says it all. Thanks for your time and efforts.
@simontay4851
@simontay4851 6 ай бұрын
HAVE a look, not take.
@johncoops6897
@johncoops6897 6 ай бұрын
"Take a look" (used in the USA) is the commonly used equivalent to "Have a look" (used in the UK). ... At least he didn't say *"off of"* instead of "from". 😢
Internal Resistance Tester TR1035 - Can we find a faulty cell?
20:57
Off-Grid Garage
Рет қаралды 41 М.
“Don’t stop the chances.”
00:44
ISSEI / いっせい
Рет қаралды 62 МЛН
My scorpion was taken away from me 😢
00:55
TyphoonFast 5
Рет қаралды 2,7 МЛН
UFC 310 : Рахмонов VS Мачадо Гэрри
05:00
Setanta Sports UFC
Рет қаралды 1,2 МЛН
Review/Teardown of a BSIDE U0 Clamp Meter
17:49
Kerry Wong
Рет қаралды 4,8 М.
Measuring Battery Internal Resistance
17:19
Adam Welch
Рет қаралды 56 М.
Fnirsi HRM-10 internal resistance meter - test and teardown
13:31
DiodeGoneWild
Рет қаралды 14 М.
Is it Time to Replace your Batteries? Here is how you can find out!
11:21
More Details - BT168 and BT168D Battery Testers
15:44
Julian Ilett
Рет қаралды 16 М.
Heltec Internal Resistance Tester HT-RT01. Not what I was expecting...
32:52