Probe Change Rack Calibration | CMMXYZ Tech Tips

  Рет қаралды 8,315

CMMXYZ

CMMXYZ

Күн бұрын

Probe Changers and How to Calibrate a Probe Changer in PC-DMIS.
Welcome to another video on tech tips from CMMXYZ. Today's topic is probe changer and how to calibrate a probe changer. Probe changer is a mechanical rack that can hold the detachable components of the probe like stylus holders and modules to make an automated switch to radio's programming time. In this video, the steps taken to prepare for calibration can be used for any probe changers that PC-DMIS supports.
Before we start, calibrate the required probe modules and clamp the probe changer on the CMM at a suitable position. The first step in the process is to select the probe changer. Go to edit, preferences, probe changer. In probe changer dialog box type tab, go to probe changer type and click on the dropdown. Here, you can find the list of all the probe changers that PC-DMIS supports. I am using MCR20 probe changer for TP20 with six ports for this video. Then apply to add it. If you have more than one probe changer, then enter the number of probe changers and define individual probe changers. Keep the docking speed at a default value of five millimeters.
Next, in the mount point tab, select the active probe changer and keep the wrist angle at A0, B0. Using the jog box, move the probe to a safe distance from the rack and click on read machine to populate the X, Y, and Z values. Then apply to save the changes. Then go to ports tab. In number of ports box, specify the number of ports on the probe changer. Then apply to add the ports. Next, add the modules to its respective ports according to your setup on the CMM. I have probe 2 by 13 millimeters in port 1, probe 3 by 20 millimeters in port 2, and probe 3 by 40 millimeters in port 3. Leave the rest of the ports empty and apply the changes. The module in the first port will be used to calibrate the probe changer.
Next step is calibrating the probe changer. Go to calibration tab and select the active probe changer and the active probe file that is 2 by 30 millimeters and the orientation of the probe used for calibration. That is A0, B0. Then click on calibrate to calibrate the probe changer. Then calibration starts. Select the loaded profile and press OK to rotate the probe to A0, B0 orientation. A message pops up asking you to open all the lids and remove probe from port one. Press OK and take a hit at the center of the silver front face between port one and port two. After taking the hit, click on continue and a second message pops up asking you to take a hit on top of the plate between port three and port four.
Take the hit and click on continue. Next, the third message asks you to take the third hit on the front face of port five and six. Press OK and take the hit. Then when we click on continue, the CMM switches to DCC mode and calibrates the probe changer.
After calibration, go to ports tab and check the location of the ports. The Z value should be same or close to each other. Then press OK to save the calibration data. To switch the probe using probe changer, load the profile and go to operations. Load active probe. This will automate the switching of the probe modules. If the modules are not seated properly after the switch, we'll have to adjust the modules one time after the pickup. And the module's position will be corrected during the drop-off. This was a video on probe changer calibration. Thank you for watching the video.
CMMXYZ - Coordinate Measuring Machines - www.cmmxyz.com/
Follow us on Facebook - / cmmxyz
Get the latest on Twitter - / cmmxyz
Follow CMMXYZ on LinkedIn - / canadian-measurement--... .
Learn more about CMMXYZ on our blog - blog.cmmxyz.com/blog
Subscribe to our KZbin Channel - / cmmxyzvideo

Пікірлер
Polyworks CMM Probe Building | CMMXYZ Tech Tips
4:29
CMMXYZ
Рет қаралды 1,6 М.
PC-DMIS ClearanceCube | PC-DMIS Tech Tips - CMMXYZ
8:06
Alex hid in the closet #shorts
00:14
Mihdens
Рет қаралды 17 МЛН
UNO!
00:18
БРУНО
Рет қаралды 1,8 МЛН
Sigma Kid Hair #funny #sigma #comedy
00:33
CRAZY GREAPA
Рет қаралды 39 МЛН
БАБУШКИН КОМПОТ В СОЛО
00:23
⚡️КАН АНДРЕЙ⚡️
Рет қаралды 18 МЛН
PC-DMIS Read Point Alignment | PC-DMIS Tech Tips - CMMXYZ
6:39
Romer Arm Demonstration
7:48
Les Sinnock
Рет қаралды 15 М.
How-To: Align Tool Rack on ZEISS CMM with VAST XXT Sensor
5:26
Inspection Engineering
Рет қаралды 11 М.
PC-DMIS Star Probe Calibration | CMM Tech Tips - CMMXYZ
5:25
How to EASILY transform coordinate system in PC-DMIS
10:13
Measure in 3D
Рет қаралды 6 М.
PC-Dmis Probe Calibration
16:28
Automettech
Рет қаралды 6 М.
Influence of Temperature on Measurement - Metrology Training Lab
11:10
Mitutoyo America Corporation
Рет қаралды 18 М.
Cleaning the CMM probe system
1:29
Tõnu Leemet
Рет қаралды 1,5 М.
PCDMIS - Import CAD & Transform
11:34
Brad Young
Рет қаралды 7 М.
Look, this is the 97th generation of the phone?
0:13
Edcers
Рет қаралды 7 МЛН
iPhone 15 Pro в реальной жизни
24:07
HUDAKOV
Рет қаралды 484 М.