can this be used to test inhert oxygen spaces less than 2 percent?
@drstevengoodman179 Жыл бұрын
Unfortunately not. For those levels you need to use a paramagnetic oxygen sensor
@HarbingerDOOM6 жыл бұрын
I ran a zero calibration and my values started skyrocketing and becoming unstable after. Any idea why? Thanks.
@drstevengoodman1796 жыл бұрын
HarbingerDOOM, did you also run a span calibration? When values start sky rocketing it normally indicates that the signal generated by the zero and span are too similar, most often due to a low UV output PID lamp. So most likely your lamp is on its last legs.