Mindray SV300 Oxygen Supply Failure Error

  Рет қаралды 183

ROYSANTIK VLOG

ROYSANTIK VLOG

Күн бұрын

Пікірлер: 8
@Pramod_Revolution
@Pramod_Revolution 20 күн бұрын
02 fails that mean of time I changed the o2 sensor .after 2 weeks its get the same error while self check
@roysantik
@roysantik 20 күн бұрын
Preplace proportional valve n0t the O2 cell..
@Pramod_Revolution
@Pramod_Revolution 20 күн бұрын
Sir may i know the name of the spare .which was changed
@roysantik
@roysantik 20 күн бұрын
This is O2 proportional valve sir if you have Oxygen supply Failure this part is defective. replace it
@daisizhou
@daisizhou 23 күн бұрын
Hallo .Why does the SV300 proportional valve always break down? Is it a parts defect?
@roysantik
@roysantik 21 күн бұрын
Yes it was upgraded already to 120psi before it was only 60psi so they upgrade to 120psi, Mindray increase the capacity of the holding pressure of proportional valve. mostly it was damage due to the high pressure. The standard pressure in the manifold in the hospital was 60psi or 4Bar, sometimes if the manifold PRV has a leaked the pressure will increase so that's the cause why the proportional was damage.
@Pramod_Revolution
@Pramod_Revolution 20 күн бұрын
Sir may I know the name of the spare which was changed
@roysantik
@roysantik 20 күн бұрын
@@Pramod_Revolution O2 Proportional valve sir.
Patient Ready Valleylab™ Medtronic™ FT10
2:28
ESU MEDICAL SOLUTIONS
Рет қаралды 70
GE Corometrics 170 series Fetal monitor
6:22
The_Biomed
Рет қаралды 56
Installing Mindray Central monitoring System
50:25
ROYSANTIK VLOG
Рет қаралды 138
Quanta Holmium Laser Low Flow Error replaced Water Filter
17:02
ROYSANTIK VLOG
Рет қаралды 78
Russia expelled from the military base / A stab in the back
12:49
Engine Sensors - Basics. 3D Animation
16:19
CARinfo3d (En)
Рет қаралды 963 М.
Raspberry Pi 500: Best Linux Keyboard PC?
23:23
ExplainingComputers
Рет қаралды 72 М.