Sir filter integrity ke bare me video banayiye na plz
@Vikayadavbpharmhhhh7 ай бұрын
Difference between lag time and equilibrium time
@ravaljinang83897 ай бұрын
Diff. Btwn lag time and equlabrium time
@sumantakumarjena4845 ай бұрын
Sir,Sip ka Impact/Risk assessment ka video.
@rakeshbariya-v7zАй бұрын
mfg CPP ka video bhejo
@domendrasahu40582 ай бұрын
Sir psg ka pressure kitna rhna chahiye
@mayankdhiroliya7 ай бұрын
Sir, Gas filter or J tube ka kya use hota hai storage vessel me
@PHARMAVEN7 ай бұрын
Gas filter for filtering gas when solution level comes down J tube is to transfer solution from mixing vessel to filtration vessel
@namratajain4152 ай бұрын
During operation When temperature drop on 121 so in case sip cycle valid or not
@PHARMAVEN2 ай бұрын
It cannot be considered as valid but need more information to understand that at what time it drop, for how many seconds or minutes, what was total time of sterilization and at what location it went low and at how many locations it went low
@Rakesh-wj9vn6 ай бұрын
Sir in vessel line there is formation of condensed due to which temperature not reaching 121 so why this condenset is forming in lines what is the reason?
@PHARMAVEN6 ай бұрын
I will make video on this Can you share vessel size , steam trap is there or not, vessel line diameter, length of line, how sensor is placed? Give as much of information available with you as you can
@Rakesh-wj9vn5 ай бұрын
500L vessel yes steam trap is there off inch diameter sensor is placed horizontal to the line Line length is not sure may be 3/4 feets
@mohitbisht20414 ай бұрын
Sir what is the leak rate value in sip ... In our company the acceptance limit 0.20 bar that is higher then autoclave why ??
@PHARMAVEN4 ай бұрын
In SIP, usual practice is to set 1 bar pressure and 0.1 bar reduction is allowed after 30 minutes of hold. Unlike autoclave, in SIP, pressure is maintained after SIP throughout before filtration is started hence there is no ingress of non-sterile air in vessel.