kzbin.info/www/bejne/j3Tbd59_i9t9fLM How To Develop Terminal Sterilization Cycle? Why Air Removal not required in Terminal Sterilization? Why Vacuum not required in Terminal Sterilization? What is D Value? What is F0 Value based Sterilization Cycle?
@kamalkishorpant9720 Жыл бұрын
Many many thanks Sir for teaching , gratitude 🙏
@nitinkmahajanmahajan2237 ай бұрын
Nice video
@bhupendrashukla1972 Жыл бұрын
Useful
@nirmalsartana4061 Жыл бұрын
Very nice sir
@tarunthakur2009 Жыл бұрын
Many thanks sir
@PHARMAVEN Жыл бұрын
Thank You For Your Appreciation, Kindly also share to your friends and groups and juniors 🎉🙏👍
@pukhrajmaske7047 Жыл бұрын
Sir hindi me sabhi video bano
@Helmet_red Жыл бұрын
Sir could you please explain about the ventilator cycle in brief please
@BrandBabuSoni Жыл бұрын
Hindi me bhi bana do ek video
@Animal-m5y9 ай бұрын
How to detrmine fo value using bacilus subtilus at 108 degree for 60 min
@ManishKumar-fp3wn4 ай бұрын
Which guidelines refer terminal sterilization.???
@kalyanvenkatesh9523 Жыл бұрын
F0 value development?
@md.shahidul9013 Жыл бұрын
Nice,My question,what’s area is required for terminal sterilizer,is non class area okay? Please give answers with reference
@PHARMAVEN Жыл бұрын
After filling and sealing, the product is basically finished product and doesn't need to store in classified area. For terminal stabilizer, CNC (Controlled Non Classified) area is acceptable
@niravbhatt1473 Жыл бұрын
Thanks sir Temperature mapping required in what size containers?
@PHARMAVEN Жыл бұрын
100 ml and above
@marwanfahmy8608 Жыл бұрын
If i use over kill approach of 8log reduction why i consider 121/10min cycle not a Sterilization cycle.and what is the minimum time parameters if I need to change for higher temperature as 134.knowing that f0 is12.... thanks.
@PHARMAVEN Жыл бұрын
Overkill depends upon D value of product, for example if D value is 1, then 121 degree C at 10 minutes will give 10 log reduction, Overkill means 12 log reduction.
@AnilSingh-hg5pz Жыл бұрын
How vials are kept inverted or upright during ts
@PHARMAVEN Жыл бұрын
Upright only during terminal Sterilization
@bholuyadav9276 Жыл бұрын
Which guidelines is for terminal sterilization please reply
@PHARMAVEN Жыл бұрын
European guide: Guideline on the Sterilisation of the medicinal product, active substance, excipient and primary container USFDA: Guidance for Industry Submission of Documentation in Applications for Parametric Release of Human and Veterinary Drug Products Terminally Sterilized By Moist Heat Process