Ini saya ada do meter ysi 20 pro. Tampilan menjadi 0 d display kenapa ya
@nanadaime48243 жыл бұрын
Do saya over . Dan tidak bisa kembali normal 😭😭😭😭
@devysofiasari37734 жыл бұрын
Kalo persentasenya lebih dari 100% gimana ya kalo di kalibrasi menjadi over
@arachemmalaysia18494 жыл бұрын
Dear Devy, you have to make sure the meter is turn on within 10-15min before calibration is start. The condition of the sensor, have to keep it in moist sleeve during this waiting. One more thing, you have to make sure membrane in a good condition, and KCl already fill up in side the membrane and no air bubbles present. Hope, this is help. :)
@nanadaime48243 жыл бұрын
Dan tombol nya mengkalibrasi sendiri 😭😭😭
@gd-mz1xv Жыл бұрын
Saya ada DO meter YSI Pro 20 tombol On/Of tidak berfungsi kenapa ya?. Setelah Caliberasi succes, terus semua tombol macet. Bagaimana mengatasinya? Terimakasih.
@gd-mz1xv Жыл бұрын
Bagaimana cara komunikasi dengan technical support untuk konsultasi alat YSI Pro 20
@eliapipitrahmawati45943 жыл бұрын
Sangat Membantu 🙏
@frmnmhmds1475 Жыл бұрын
Kalo persennya dibawah 100% bagaimana mengatasinya?
@CLOWZY992 жыл бұрын
Apa bisa kalibrasi tanpa Spon? Kalau spon hilang apa ada cara lain terimaksih
@duniaauto59205 жыл бұрын
Klau do meter tiada paparan suhu dah DO ap punca nya ya?
@arachemmalaysia18494 жыл бұрын
Afraid the meter/ cable/ sensor is spoiled :) You need to swap with good condition meter/cable/sensor to troubleshoot this.
@swarthazine29264 жыл бұрын
Calibrasi under kenpa ya?
@arachemmalaysia18494 жыл бұрын
Boleh cuba untuk ganti elektrolit ATAU membrane ATAU mengilapkan silver katod dan cuba kalibrasi semula. Pastikan meter sudah dihidupkan lebih kurang 10-15 minit sebelum mulakan kalibrasi. Jika gagal, berkemungkinan probe rosak.