Could you please explain in greater detail why the vent senses a leak on the flexitrunk interface but not the RAM interface. Doesn’t it sense that there is missing flow in either case, at the expiratory end of the vent?
@samroller5788
@samroller5788 2 ай бұрын
So what happens if the baby doesnt have a pacifier in? At this point we would use a chin strap to prevent a leak from the mouth? I wasnt aware an open mouth can cause a leak
@sternbildmedical1367
@sternbildmedical1367 2 ай бұрын
@@samroller5788 the mouth is a potential leak. In adults, bipap masks can be used to stop mouth leaks and maintain CPAP. That option doesn’t exist in the NICU so pacifiers or chin straps become and option. Unfortunately, cannula prongs are very far from the expiratory limb. If a baby needs to exhale CO2, they have to do so around the prongs or out of the mouth. In a true NCPAP interface, the nose of the baby is very close to the expiratory limb making it possible to exhale out of the nose without increasing the work of breathing. You can keep the mouth closed with a pacifier and baby can still safely exhale CO2
@samroller5788
@samroller5788 2 ай бұрын
Peep to 25??
@sternbildmedical1367
@sternbildmedical1367 2 ай бұрын
@@samroller5788 if I said peep of 25 I meant pip of 25 and peep of 5. I speak faster than my brain sometimes.
@nickl.7305
@nickl.7305 2 ай бұрын
Hi! At 32:48 you demonstrate blue RAM cannula at 60% occlusion but the screen is washed out in the video. How high is the pressure with this 60% occlusion at 6LPM? (Thanks for the very informative video)
@sternbildmedical1367
@sternbildmedical1367 2 ай бұрын
@@nickl.7305 when I set the bubble CPAP to 5 cmH2O, the green RAM consistently reads about 2 cmH2O and the blue (which is bigger with more occlusions) reads 3 cmH2O.
@espoir8131
@espoir8131 10 ай бұрын
This was super helpful. Thank you. I was looking up how to check the maximum pressure relief (“pop off”).
@jodyjohnson4903
@jodyjohnson4903 11 ай бұрын
Great info. This is really helpful for learning how to adjust the FP system.
@jodyjohnson4903
@jodyjohnson4903 11 ай бұрын
Great illustration of discrepancy of set and delivered and measured pressure. This ventilator can be calibrated with prongs and pressure at the interface (PEEP if) measured as well. Delivery of pressure with the RAM is very low.