Can you explain what is and how to do "debranching"
@MSK-s7w3 жыл бұрын
Amazing and great case !! We should always pay attention to pressure before injecting contrast.
@shangz02163 жыл бұрын
Thanks for the educative case sharing.
@ramidudin13 жыл бұрын
Thank you for share this case, I think conclusion to secure distal wire before any inflation balloon or sent. Another point I will stop and avoid further contrast injection to avoid dissection expanding to right coronary sinus and ascending aorta because patient stable and she has good collateral from left.
@arminmarashizadeh55402 жыл бұрын
Thank professor for your educative case as usual, can you explain more about debranching ? How did you reentered?
@praveenalane43313 жыл бұрын
Thanx for the case professor..what alternative wires do you suggest other than gladius for debranchibg in star approach ?
@manosbrilakis3 жыл бұрын
Polymer jacketed guidewires, such as Fielder XT or Pilot 200.
@denizaktürk-j7n3 жыл бұрын
Thank you very much for shareing this case. I would like to ask what if you had another complication at septal artery that was keeping the patient stable?
@manosbrilakis3 жыл бұрын
A problem with a septal branch would likely not be a big problem, but compromising LAD flow could be fatal. Using a safety wire and meticulous attention to ACT and pressure waveform are important to prevent donor vessel injury/thrombosis.
@denizaktürk-j7n3 жыл бұрын
@@manosbrilakis Thank you for your kind response. It is a great honor for me.