Fluke 179 is about 10 deg high also. Thanks for passing on your knowledge GFM.
@grayfurnaceman7 жыл бұрын
Yeah, I went after one of their engineers about it and got nowhere. GFM
@rob440sixАй бұрын
Maybe they've fixed the issue in the intervening years because I just checked two different K-type probes in a crushed ice bath on my Fluke 179 - one read 0.1ºC, the other -0.1ºC, but in ten minutes it had settled to 0.0ºC. Thanks for the tip about the crushed ice instead of cubed, GFM.