Sir The exhaust control not available in the market
@EstherKirabo-u6c2 ай бұрын
My Arm type BP monitor is not giving diastole and pulse rate readings with very high systole readings and then no alarm is heard ,what could be the cause
@dr.jayprakashbharati2196Ай бұрын
Screen not showing digits properly
@Mr0117962 жыл бұрын
How to get pcb board to replace citizen bp 432b model to get repaired.
@MiguelFloresV.3 жыл бұрын
I have the ch-453 version, when I press start it shows "--- ---" and if I press it again it turns off. What should I do?
@Vktyo3 жыл бұрын
If you got solution plz mention the solution
@psoberoi34702 жыл бұрын
My CH 432 when switched on shows six horizontal dashes like - - - - - - and also does not pump air nor changes its dashes. How do I correct and also is it possible for me correct or have to contact a technician.
@muhammadsaqibghauri7437 Жыл бұрын
I have citizen BP machine but during using after 3 years I feel some extra tightness during BP check on under cuff .
@mr.thumbsup27422 жыл бұрын
Good day friend, how can I repair my ALP K2 digital BP monitor that seems to be low on pressure and cannot complete BP monitoring results.
@sudhirthingalaya49673 жыл бұрын
My ch432 citizen monitor is showing err1 what could be the problem
@sudhirgajananprabhune7432 Жыл бұрын
Where will i get ear plug in mumbai
@mariomontano3984 жыл бұрын
Bello greatings I hace a cuestión my tensiometer ch453 si now wit the screen blank after conect 9volts in the dc input. And the Pump not stop and star Lomelí. What can i go?
@peterjensen4462 Жыл бұрын
LOVE your voice 🤖🤖🤖🤖
@Banushrye14 жыл бұрын
My BP monitor. Power switch is not functioning properly.what SHOULD I do?
@shamsheeralam85463 жыл бұрын
Hi, is there any way to replace batteries from mobile charger?. I am tired of changing batteries again n again. please make a video on setting a mobile charger or adapter in place of batteries in this citizen automatic blood pressure machine.
@Banushrye14 жыл бұрын
I live murshidabad of West Bengal.pls let me know nearby service centre.
@nomanshahid7179 Жыл бұрын
I have microlife digital blood pressure monitor model BP 3AR1.3P having an air loses issues unit 38, 40 but not increase and exhaust loses the air
@shivkumarswamy2573 Жыл бұрын
I have Dr trust bp monitor machine, If inflation starts, it doesn't stop till it reaches 270 or 290, how to fix this problem....
@asadpatel7773 жыл бұрын
My machine continues to blow, it's hurting alot
@nenetcomia94993 жыл бұрын
I like your discussion abt . Blood pressure monitor. What about if my equipment bp monitor. With same model and brand is not functioning for my satisfaction. Whenever I use it lately it sqeezes my arm so much. It is painful indeed and now I cant use it anymore. Thank you for any answer I might receive
@brandstore357810 ай бұрын
this is bp monitor and it will inflate the cuff .. if u think its squeezing too much.. it means u hav worn the cuff too tightly.. the proper way is to keep one finger or less looseness when u wear the cuff.. do not wear the cuff too tight .. bcoz it has to inflate...
@ricardoobandoceciliano84064 жыл бұрын
hello , what can be the problem if the numbers keep going up but it doesn't stop and it doesn't give the pressure result too
@muhammadnomanyounus83735 жыл бұрын
Dear sir I have the model ch 456 and when starts it working and when teached at 90 it's show error 3.what does it's means and how to resolved this error?
@cesaryusti23873 жыл бұрын
Hello brhother, you can make un video, were explain the components electronic de tensiomets, please
@mahsharsiddiqui66123 жыл бұрын
Monitor is black how to correct
@KingBravo-lo3vc10 ай бұрын
My problem is the start button doesn’t start the pump, but the memory button does. And the only way it stops is by disconnecting the battery.
@jafarvakili91086 жыл бұрын
, Thanks for your good and useable video, i have a wrist Omron blood pressure monitor, but numbers doesn’t come on completely, plz let me know what’s the problem? Unit isn’t available now to read the model,but its in series 652.
@lailaceleste6 жыл бұрын
Dear Jafar. Thank you for your comment. We are CITIZEN so we can not give you advise about Omron items. Please contact their service center in your country.
@binhminh4175 жыл бұрын
Great video! I have a problem with the Medline MDS4001 Automatic Digital Blood Pressure. It worked fine with batteries until I switched to an AC adapter. Unfortunately, the unit was not working with the AC adapter so I switched back to battery. The unit became malfunctioned (the start/stop buttons didn't work & the screen monitor went completely blank) even with new batteries. Appreciate your input.
@islandgate-igt-citizenheal43635 жыл бұрын
Thank you very much for the comment. Regarding your BPM we can not give you a precise answer because is not CITIZEN brand, you should contact your local dealer. But as per your description, this must be a failure of the electric circuit. Some brands don't take care of this point and align the 2 power source systems correctly, so when you go from batteries to AC, and come back to batteries then the system doesn't respond. Our advise is to send it to your nearest repair center from your BPM brand. Hope the answer is useful.
@abowden5567 жыл бұрын
It's nice to see well, ANY electronics manufacturer doing repair tutorials. I guess things are different in the medical industry...
@islandgate-igt-citizenheal43637 жыл бұрын
Dear abowden556, thank you for your comment! From CITIZEN we try our best to be transparent and honest with our customers, so following this trend, why not showing the functioning of our devices?! :) We are proud of them and if there is a problem we are here to find the best solution! Plus, a little learning of how a BPM works is always useful to understand little more how the machines work. Thanks again for the encouragement!
@ashishchandrawanshi96526 жыл бұрын
Sir i want to know that whats the name of the thing which pump the air its a part on a motor.
@ledi6664 жыл бұрын
Hello sir ...my device just went blank after plugging it to an adaptor...please help
@welcometomyworld....5786 Жыл бұрын
Good job ❤❤
@eiouldelarosa83685 ай бұрын
How about error1
@yogendersingh93763 жыл бұрын
My machine is showing E 1 What's wrong nd how it should be repaired tell me please
@justinkf29084 жыл бұрын
Sir l am using enbee make model NB 103N . It's all ways shows high reading. Valve, Motor are ok. Calibration screw not provided inside. Some time it's deflated automatically then start again. What is the problem? Help me.
@muzaffarmehmoodyahya68173 жыл бұрын
Hello sir good informative video , if gadget take too much time tell the results, even scilent . What could be the fault. Kindly guide me.
@alepod8280 Жыл бұрын
Thank you, very helpful 🎉 success
@syedshafiq22445 жыл бұрын
dear my ch452 deflation is very slow the arm become tired and painful due to slow air release.what to do plz tell me soon.
@islandgate-igt-citizenheal43635 жыл бұрын
Dear Mr. Shafiq, thank you for contacting us. The deflation timing can not be faster because CITIZEN blood pressure monitor is still measuring while deflating. This means that it takes 3 measurements, while inflating, once is fully inflated, and while deflating, and the final result is the average of these 3 measurements. To avoid having pain in the arm please try to wrap the cuff not so tight. When you wrap the cuff around your arm don't press it too much, just tight it enough to not fall down from the arm, and this way the pressure will not be too high to hurt your arm.
@arunavabhattacharyya34377 жыл бұрын
where is the sensor which senses blood pressure...?
@قناتللمنوعات-ي8ل3 жыл бұрын
Thank you so much verry interesstin video.
@vijaykumar-by3hn4 жыл бұрын
Tell replace compressor system
@medicalali7815 жыл бұрын
Dear I required citizen bp spare parts in Pakistan .how it possible
@usmanbashir6695 жыл бұрын
Sir my device stuck on any digit on pumping goes countinues and after some time give error 2, where to find new cuff, waiting for your reply
@islandgate-igt-citizenheal43635 жыл бұрын
Dear Sir, thank you for contacting us. For new cuff you have to request from the local supplier. Usually the shop where you purchase the device will be able to tell you who is the distributor in your country.
@SuperGaguu5 жыл бұрын
Sir Please tell where I can get spares in India of this BPM as I want solonoid valve & it's transistor
@islandgate-igt-citizenheal43635 жыл бұрын
Dear sir, thank you for contacting us. For India territory you can contact directly the local distributor. Please follow this link: www.meharhealthcare.com/
@shahidsaeed8039 Жыл бұрын
My machine displays err 1
@shubhamkeshari14055 жыл бұрын
Sir my BP machine shows ---- when I start it...please give me solution.
@islandgate-igt-citizenheal43635 жыл бұрын
Dear Sir, thank you for contacting us. the lines you see means that there is no memory recorded, so you have to place the cuff in your arm and then press the start button. by doing this the cuff will start inflating and your pressure will be measured. If this suggestion doesn't solve your problem please let me know the model number and we will review the switch ON procedure again.
@shubhamkeshari14055 жыл бұрын
@@islandgate-igt-citizenheal4363 The model number is CH-432...the BP machine was in working condition about two years,But now has the problem of displaying ----
@islandgate-igt-citizenheal43635 жыл бұрын
@@shubhamkeshari1405 Thank you for the clarification. I consulted with our technicians and if the item was working previously (even considering 2 years of work), then most probably the problem in in the main board, which is not possible to repair.
@gokuvegeta33615 жыл бұрын
Thank you.👍
@sibgasheikh82164 жыл бұрын
Asalam o alaikum bhai
@Infartility6 жыл бұрын
Sir my machine shows - - - - - - when start . pl give me solution
@islandgate-igt-citizenheal43636 жыл бұрын
Hello and thanks for contacting us. Please note that when using the device for the first time and pressing the M button the machine will show --- --- because there are no saved measurements. When the device is OFF you need to fix the cuff in your arm and then press directly start (not M button) so the machine will inflate the cuff and measure the Blood Pressure. If the problem persist, please write to me at L.M@island-gate.com so we can give you clear instructions.
@dptzahir3 жыл бұрын
Thanks for good vdo
@abdulwahabparkar92596 жыл бұрын
i am in India I have a CIGAHealthcare UK machine which is comimg on but further no actions please guide me to repair it
@lailaceleste6 жыл бұрын
Dear Mr. Abdul, thank you for contacting us. Note that we are CITIZEN devices and we don't have information about CIGAHealthcare brand. Kindly contact their main office or their appointed repairing service in your country.
@pawankailash48914 жыл бұрын
Sir my BP machine is showing error 1
@pawankailash48914 жыл бұрын
What is problem sir please saying about
@rizkhanna75503 жыл бұрын
perfectly explained....
@jhuneagravio94605 жыл бұрын
hello sir/mam we had CH--453 long time we dont use it now i try to use it again but not working properly i already change the battery..it show error 1 it stop in 70 SYS and not continue in DIA where i can bring the machine for repair? im in Calamba Laguna area Philippines...thank you
@islandgate-igt-citizenheal43635 жыл бұрын
Dear Mr. Agravio, sorry for the late reply. Error 1 means air leaking, so according to your explanation, and considering the long time without use of your BPM, the most probable case is that the interior pipes got dry and cracked (allowing the air to not press correctly), or there is some puncture in the cuff, making not possible to inflate correctly. For repair in Phillipines you can contact directly our Japan office at sales-oe@systems.citizen.co.jp , and they will provide information for an authorized repair shop in your area.
@marda7525 жыл бұрын
just buy a new one, they cost less than 30 bucks
@wojtekzieba68353 жыл бұрын
I'm in Canada and the cheapest one cost 140 dol, there are some cheaper ones but reviews are like 2 stars.