Gracias gran video, ingeniero tendrás alguno de mantenimiento a un advia560 de Siemens gracias de nuevo
@tanguykpofoessi80742 жыл бұрын
I would like to know something sir What to do if the results do not give despite checking the lamp and filters which are normal
@esaugalvez2 жыл бұрын
I don’t quite understand. That’s a bit general. If you don’t get results. Even when everything seems to be normal. But how are results, is there reproducibility in the results ?
@tanguykpofoessi80742 жыл бұрын
Good morning Sir I'm biomédical Technicians Can you help me to have a service manuel of this device I have erreur 9 when i start the équipements
@esaugalvez2 жыл бұрын
Disassemble the pump. Clean it, lubricate it and make sure the pap motor is in good working condition. Assemble everything again. And it must work again
@tanguykpofoessi80742 жыл бұрын
@@esaugalvez yes I disconnected the pump and cleaned everything before reassembling The longer the problem persists... it's like the engine won't even start I want Also know if the lithium batterie 3.6v conditionne the fonctionnement of this motor!?
@carlosignaciosantanazapien6032 жыл бұрын
Se pueden limpiar los filtros del spinlab cuando ya tienen hongo ?
@esaugalvez2 жыл бұрын
No se puede. Tienen un tiempo de vida. Se degrada el material de las capas y no hay de otra más que cambiarlos
@carlosmuro2605 Жыл бұрын
Dónde puedo conseguir los filtros , los cristales y la lámpara
@thosynlee6833 Жыл бұрын
@tanguykpofoessi8074 Please I'm having error code 10 on my machine. It's not completing the initialization. I noticed the filter wheel is not rotating during initialization. Could that be the problem?
@esaugalvez Жыл бұрын
Error 10 refers to Filter Wheel. It has to move freely, If it doesn’t move freely or doesn’t move at all, then stepper motor has to be checked.