Hi. I would like to request which button you pressed to make zero initially.(at 1:17)
@theacez90322 жыл бұрын
Clear (zero) bottom right next to the enter
@joshuapesina27094 жыл бұрын
Do you have a pdf written procedure for this
@fluke4 жыл бұрын
Hello Joshua - We do offer some similar Application Notes that you can check out on our website. Try this link: bit.ly/3drUYpN This is our HART Calibration app note with examples and procedures. Hope this helps!
@sebastianamado77585 жыл бұрын
Good afternoon, an innocent question. What error percentage does the calinter indicate? in 0 PSI try relative error (Vm-V) / V and I can not match that error indicated by the video.
@fluke5 жыл бұрын
Hi Sebastian - Our Product Specialists would be happy to help troubleshoot. Please give us a call at 1-800-443-5853 or email at fluke-info@fluke.com. Thanks!
@johntrauger684 жыл бұрын
I'm doing a three-point calibration on a 3051 Rosemount 0-25" h2O with a Fluke 754 in Hart mode. When applying 12.5" at mid, the signal back always exceeds the 12 ma I expect. Usually around 15 ma. It's acting like the sensor output is skewed at the mid-point only. Any suggestions?
@fluke4 жыл бұрын
Hello John - It sounds like you might have the calibration set for square root. It should be set for linear. Let me know if this works out. Please feel free to reach out again or contact our Technical Support team at 1-800-443-5853.