Hola, los GX-8000 se calibran con butano o con propano, porque hay está diferencia ??
@Yangyuzhitian2 ай бұрын
👍
@georgeshmelev6893 жыл бұрын
Good explanation. Thks man👍🏻
@juanclaros91404 жыл бұрын
Hello, the GX-8000 are calibrated with butane or propane, why is there a difference?
@juanclaros91404 жыл бұрын
How does the calibration expiry time change??
@Jkungames4 жыл бұрын
Only connecting with the software from riken keiki . The maximum expire due is 365 days
@juanclaros91404 жыл бұрын
@@Jkungames Hello, the GX-8000 are calibrated with butane or propane, why is there a difference?
@Jkungames4 жыл бұрын
@@juanclaros9140 hi! My specially area is gx-2009. O haven't work with gx-8000 but I believe you can calibrate It setting the gas measured. You can set what gas is measured by accessing the "gas comb " menu . In gx-2009 you can measure the explosive gas with CH4, HC or lower explosivity limit. By default, ALL the equipments I ever calibrated was configured as LEL porcentage. You can check your gas cylinder to know the gas to be measured. I only work with gas mixtures with 2,5 % of CH4 ( 50 % LEL )
@mozonlab21282 жыл бұрын
Dear Sir , Could you pls tell me the Fail or Pass showing automatically or need to something ? and Like Bw Honeywell Gas Detector showing Cal Date and Cal Due 180days or 365 days but here is not showing anything ? Is it true ?
@kos67794 жыл бұрын
Is the password same for all the same models?
@Jkungames4 жыл бұрын
For the aversge equipments. Riken keiki CX-5 for example, has other passwords .