I noticed that you used an Isobutane ( C4H10 / 0.9% ) mixture for your 50% LEL range. But the instrument is using a CH4 sensor ( as shown at the start-up screen ). Although you can indeed indicate 50% lel during the calibration sequence, is this not a little erroneous for that CH4 sensor? Shouldn’t the calibration gas mixture match the sensor being tested? Apologies if this seems like a daft question! 😊