Can you upload a video on how to calibrate the analyzer?
@MedicalLabLadyGill6 жыл бұрын
Anthony Dzrekey I won't be doing that for a while but will try to make one when I do!
@MedicalLabLadyGill6 жыл бұрын
Anthony Dzrekey I apologize for not having posted your request. We are currently waiting to have a service representative come and take a look at our machine. I have been unable to do calibration due to some issues.
@hamzabeldjouzi7860 Жыл бұрын
Le mode ďanalyse actuel ne correspond pas à la position du tube.changer le mode ou ajuster la position du tube
@taofeekoladosu59362 жыл бұрын
Thanks for video , but how can i reach you if want to learn more about the machnie
@MedicalLabLadyGill2 жыл бұрын
Thank you. I no longer use this machine and it is not supported by the company.
@ArifKhan-gy8st2 жыл бұрын
Madam i hve problem, im using mindray bc 3200, showing Vacuum Filter error, what i do i changed the small filter already, problem still there
@d.i.yrepairnmaintenance9538 Жыл бұрын
Great video..
@medicalinstruments15934 жыл бұрын
nice video
@ranjushakp10783 жыл бұрын
How to remove hb errors in voltage prblms?
@LearnEnglishMaster-8042 жыл бұрын
Please help me i have cbc analyzer urit and i have a problem that give error and no giving result
@Yasmina1119-r1y2 жыл бұрын
Hello, I have a problem that she writes to me agrégat hgb, please, I want a solution
@kacemkacemo77422 жыл бұрын
I'm facing the same problem. Did you find a solution. If so please tell me.
@abdelmageedahmeed85515 жыл бұрын
What the mid in WBC differential refer to?@Medical Lab Lady Gill
@medicallabtechnologistrahu18653 жыл бұрын
so nice video
@hassanahmed9105 Жыл бұрын
Can i used 70% alcohol for cleaner to replace probe cleaner
@jeanlouise65575 жыл бұрын
Hi! I just wanna ask, if I am going to use a probe cleanser, should I feed it as a sample as well? I hope you reply. Thank you!
@MedicalLabLadyGill5 жыл бұрын
There is a specific cleaning mode to run probe cleanser (probe cleanser cleaning). An often too common problem with this analyzer is that the internal fluidics do not get cleaned appropriately as often as they should and a internal clot forms. There is a section in the manual where it talks about zapping and flushing the apperatures 4 times but this protocol is embedded in the pages of the manual and not readily found. Our Mindray analyzer is no longer in use because it did get an internal clot and the company jo longer supports the analyzer. We now have a Sysmex XN-550.
@hassanahmed9105 Жыл бұрын
Can i used for 70%alcohol as a cleaner cleaning
@kamleshtiwari46274 жыл бұрын
hii i have mindray 2800 cellcounter & it shows me syringe motor error n vaccum low pls suggest the solution for my problem..
@MedicalLabLadyGill4 жыл бұрын
I am sorry. We do not use that analyzer anymore and the BC 3200 is no longer supported by the company. That being said, I will give it a shot. You want to open up the side door and look to see if the syringe (connected to tubing) is bringing up fluid as it moves. If there are bubbles you may have a leak in the tubing, clot in the RBC measuring chamber, or clot in the WBC Lyse (in door on other side of machine). Check around your analyzer for leaks from un fastened tubing. Run saline as a patient and watch to see if the cell counts go to 0. Also watch the WBC Lyse reagent tubing to see if there are bubbles and fluid moving through the analyzer. If the not cell count is not 0, zap and flush the apperatures 4x. If fluid isn't moving in the WBC lyse tubing send a needle through the straw and tubing with warm water to clear residue build up. Usually the instrument will say syringe motor error because the fluid is not moving properly through the analyzer.
@chefdzta6 жыл бұрын
Thanks for the video,
@YorkThompson4 жыл бұрын
Thank you. 👍
@ilyas47p4 жыл бұрын
What is this?
@खरीशिक्षाओंकेसाथ5 жыл бұрын
Please can you help me ? I have mindray bc 2300 . The problem is remove the error , sytume error ,,What is the Solution please
@MedicalLabLadyGill5 жыл бұрын
I am sorry, I don't use Mindray anymore. I don't know the answer to your question. Have you tried finding a service company in your area that supports the analyzer?
@nelmedlabsuppliesinc.15122 жыл бұрын
HOW TO RESOLVE THE ERROR "LIQUID PRESSURE OVERLOADED" IN MINDRAY BC5150?
@georgeantony46006 жыл бұрын
Thank you
@رحمهمحمداحمدرحمه-ف2ت Жыл бұрын
helo lm using mindary BC3000,we have problems back round errors, all analysis from wbcs, hb normal, others parameters RCS, mcv, mhcv, plt abnormal some time read stars (***) some time high reading please help me
@aysimatosun388610 ай бұрын
Çözüm bulabildiniz mi ? Aynı sorun bende de var.
@Agent-xy3vp2 жыл бұрын
I need mindray BC 2300 lcd display
@djoukham52945 жыл бұрын
Please can you help me? I have mindray bc 3000 plus .The problem is ///erreur de filtre sous vide//vacuum filter erroror //What is the solution please
@MedicalLabLadyGill5 жыл бұрын
It sounds like you either have a hose/tubing that needs to be replaced along with the filter or just the filter. There may be a something blocking the flow of fluidics through the analyzer. It is not publicized much but this analyzer needs to be cleaned A LOT.
@MedicalLabLadyGill5 жыл бұрын
We no longer have this analyzer because a clot formed and the company no longer makes parts or provides service for this analyzer. It has been a very frustrating experience. I am sorry for your situation and that it took me so long to answer. It has been a stressful spring and summer semester for me this year.
@hamzabeldjouzi7860 Жыл бұрын
ماهو حل هذه المشكلة
@gowdashashi63553 жыл бұрын
How to remove vaccine filter error
@HERAMANIGO6 жыл бұрын
Can you correct RDW parameter ? How?
@MedicalLabLadyGill6 жыл бұрын
HERAMANIGO I will need to investigate this and get back to you.
@HERAMANIGO6 жыл бұрын
I am still waiting for your response on the RDW correction. Results on all QC Levels on our BC-3200 AutoHematology analyzer were consistently High. There's nothing on the Manual Troubleshoot that can correct it. And no matter how many times you're going to calibrate , the problem remains.
@MedicalLabLadyGill6 жыл бұрын
HERAMANIGO Was the calibration successful and the QC not? Are you using Mindray Calibrator & QC? Is it mixed thoroughly before putting it on the analyzer? One thing I found out on Friday is that Mindray is/will no longer be supporting the BC 3200. You may want to start looking for another analyzer because I am not sure if they will be making the reagents much longer.
@HERAMANIGO6 жыл бұрын
The Calibration was unsuccessful ,likewise with the QC. Yes, I am using Mindray Calibrator and QC applying the principle instructions in the insert . Disappointed also to learn that Mindray ceased the production of reagents of BC 3200 'cause we just purchased the analyzer less than 2 yrs ago. Perhaps we've been decieved by the Vendor that recommended the analyzer .Why we never learned that ahead of time ?
@MedicalLabLadyGill6 жыл бұрын
HERAMANIGO I am unsure about the timeliness of the decision. They do have other models though. Maybe you can get a deal on an exchange for another model if they discontinue the reagents.
@inayatullahkakar4003 жыл бұрын
Good luck
@srajukhan6234 жыл бұрын
Platelets issue kese shi krenge
@ArifKhan-gy8st2 жыл бұрын
I am using hematology Analyser, having issue . Vacuum Filter Error, i solved by my solved, if anyone have issue like this can contact me please
@georgeantony46006 жыл бұрын
sysmex xp 100 BACK GROUND CHECK ERROR BLANK ERROR 411010.0.0 , WBC CLOG 311140.182.0 AND HGB ERROR 415000.9999.4361 SOLUTION PLEASE
@MedicalLabLadyGill6 жыл бұрын
George Antony Have you tried the Sysmex service line? They provide amazing support for their customers! I currently do not have a Sysmex analyzer and am unable to help you. : / That being said, it sounds like you have a clot somewhere on the instrument and need to thoroughly clean the tubing.
@nanuramdamor60546 жыл бұрын
Please use of Hindi language.
@soulheal63036 жыл бұрын
No need to waste this large amount of diluent because the system says it is expired, just put another bar code in diluent screen and you're good to go. Edit It's not really expired, just the system count since day you opened it rather than real day of it's expire.
@djoukham52945 жыл бұрын
ماهو الحل من فضلك.وشكرااerreur de filtre sous vide يكتبلي ميساج mindray bc 3000 plusسهل كرار.من فضلك عندي مشكل في جهاز
@soulheal63035 жыл бұрын
@@djoukham5294 Write about your problem in English please.
@djoukham52945 жыл бұрын
@@soulheal6303 Please can you help me? I have mindray bc 3000 plus .The problem is ///erreur de filtre sous vide//vacuum filter erroror //What is the solution please
@djoukham52945 жыл бұрын
Thank you
@soulheal63035 жыл бұрын
@@djoukham5294 The vacuum filter is clogged by dirt or it needs replacement. You can't fix this problem by yourself, call the authorized dealer for proper service. Best luck.