After exchanging the Expired Lyse Reagent and cleaning the Tubing it work well...thanks much..looking forward to view some more tutorials from you.
@MedicalLabLadyGill6 жыл бұрын
Dexter Takau Glad it worked out for you!
@chefdzta6 жыл бұрын
If you can show me how to run the controls and calibrator would be also of much appreciated
@MedicalLabLadyGill6 жыл бұрын
Dexter Takau I will be working on that shortly. Another viewer has made the same request as well 😀
@miglenanenkova27682 жыл бұрын
Hi! I have a problem with BC 2800, the PLTs are constantly high. So, can you help me?
@marinescug7 жыл бұрын
After you have exchanged a reagent is mandatory ro execute "prime" for that reagent. Sometimes just proble cleanser cleaning is not enough. Take out the black metallic screens from the both baths, after PC cleaning. Using a cotton swab clean with circular movements the interior of the baths. You will got a collection of fibrin and RBC debries.
@MedicalLabLadyGill7 жыл бұрын
marinescug Thank you for the excellent solution!
@abdelhakimsouidi93605 жыл бұрын
Hi, can I get your email to ask you questions about calibration and performing QC test please
@pharmhse19456 жыл бұрын
I have a problem with an automaton bc-3000 HGB = 102 g / l LOW AND RBC 7.41 HIGH
@antonisvlachavas6513 жыл бұрын
I have the same problem with the tubing.. could you explain how to unclog the lyse?? With inj water? Thank you in advanced
@shyamsinghsharma24533 жыл бұрын
What can be the reason for analyzer not displaying differential count
@dena12324 жыл бұрын
how i can connect the Mindray BC-20s with laptop or computer
@yawaboagye15534 жыл бұрын
Thanks. Please can you show us video of how to perform fresh blood calibration
@mjstudio15012 жыл бұрын
How to fix vaccume low error??
@marynonye37574 жыл бұрын
My plt does nt show, what should i do?
@kadiryaman37224 жыл бұрын
İ have a problem HGB blank voltage abnormal. How can i solve this problem?
@MedicalLabLadyGill4 жыл бұрын
I am sorry, we do not have this machine anymore.
@skshahid66054 жыл бұрын
Wash it a hypochloride
@kadiryaman37224 жыл бұрын
@@skshahid6605 Problem solved. Manual calibration done from Hgb blank voltage settings
@sbeerkaur59904 жыл бұрын
@@kadiryaman3722 call me 8194825296
@korutlanavatej13603 жыл бұрын
Clean the chamber first. Call me this number 901412983 My name is korutla sandeep
@skshahid66054 жыл бұрын
How i can remove a platelet count error a same analyser....... I give aspirate a blank sample it shows a 125 or high platelet countin blankk smaple.....
@korutlanavatej13603 жыл бұрын
First clean the chamber. Then remove in aperture Ruby Then insert the aperture. 100% clear your problem
@korutlanavatej13603 жыл бұрын
My name is : korutla sandeep My number is : 9014129183
@rollforspoons33252 жыл бұрын
Watching this for hints even though we have a totally different machine! Ours can be a real drama queen.