At 0:46 Why don't you close the high pressure block valve first instead?
@yassertantawytantawy1842 жыл бұрын
Thanks you for your video But I think the sequence to back the transmitter to service it's first open the valves of process isolation and close the equalizer in the end to prevent sell damage Because the low side have atmosphere pressure and high side well have process pressure and it's high delta pressure can cause cell damage
@InstrumentCalibration2 жыл бұрын
Yes,, You are correct
@mathewgeorge1576 Жыл бұрын
Thanks sir ji ✋
@InstrumentCalibration Жыл бұрын
Thank you George for your comment
@janetawuah25353 жыл бұрын
Thanks for your great videos...but I thought the square root extractor function is in built in the flow transmitter, so why ignore the mA readings due to the square root extractor when calibrating? Can you please explain I'm a beginner
@InstrumentCalibration3 жыл бұрын
Thanks for asking. All SMART transmitter comes with function to select square root extractor. If you do not select square root extractor at transmitter side. Then you need to select at PLC or DCS side.
@mikegrande4957Ай бұрын
Valve positions during calibration?
@abysebastian25972 жыл бұрын
Sir, how this arrangements comes practically possible. This is because LP and HP are kept close together. Since the equalizer is open, will not the air pressuse applied in HP go into the atmosphere through LP..?
@InstrumentCalibration2 жыл бұрын
Thanks for asking this question.. When we do calibration, both HP & LP Valve remains closed, and equalizing valve also remains closed. After this, open up LP vent valve, and apply pressure through HP vent valve.
@Nicofish20232 жыл бұрын
Please can you do this with a 5 way manifold showing all the steps of calibration
@InstrumentCalibration2 жыл бұрын
Hello Herman I will make the video on this topic soon
@abdulkader-lh9mj5 ай бұрын
How to get analog output ma from del p inwater column?
@mohamadsyahirsabar40972 жыл бұрын
Can u explain in details how the DCS do the calculation to get (from sqrt of mA to the flow rate in GPM). Why do we need the sqrt mA?
@InstrumentCalibration2 жыл бұрын
Hello I don't have calculation formula for DCS. Flow rate is directly proportional to square root of differential pressure. In order to convert, dpt' s output into linear, We use square root extractor in dpt.
@bino13363 жыл бұрын
For flow transmitter
@denildavis35613 жыл бұрын
Where we connect calibrator when we use 5 way manifold
@InstrumentCalibration3 жыл бұрын
You need to connect pressure calibrator to bleed valve after putting PT in service mode
@denildavis35613 жыл бұрын
@@InstrumentCalibration ok thank you sir
@InstrumentCalibration3 жыл бұрын
@@denildavis3561 You are welcome. Feel free to ask me any question
@bino13363 жыл бұрын
Without assuming Correct dp can u tell me that
@mohammadimtiyaz58242 жыл бұрын
Sir during calibration when apply pressure, low side and equalise keep open or close, plz clear my doubt
@InstrumentCalibration2 жыл бұрын
Hello Mohammad when you apply pressure, low side and equalize both valves should be in closed position.
@bino13363 жыл бұрын
Sir How to calculate the applying pressure?
@InstrumentCalibration3 жыл бұрын
It depends on your range. For example, if your dpt's calibration range is 0 to 100 inch H2O. Then divide 100 inches H2O into 5 equal part, and apply the pressure according to it.
@bobbyyoung28317 ай бұрын
Need to close the equalization valve Before pumping up. AND the pressure calibrator needs to be liquid...Air compresses. Nice video but left out a few things.
@abdulrozak5142 жыл бұрын
whats name the calibrator?
@InstrumentCalibration2 жыл бұрын
Hi dear , i would recommend use the calibrator attached in the link below. It is efficient and easy to us amzn.to/3LIy93f
@الطيرالمهاجر20123 жыл бұрын
The sequence of return it back to service is not accurate
@InstrumentCalibration3 жыл бұрын
I had made video on 3 Valve Manifolds. Please watch this video.
@hardyferrysian48132 жыл бұрын
in my humble opinion the step for calibration is 1. open the equalizing valve 2.close together high and low block valve we need eqaulizing first, if we close the low or high block valve before equalizing valve then the transmitter will damage flow transmitter have Differential range maximum base on maufacture data sheet
@InstrumentCalibration Жыл бұрын
Hello Hardy Yes, you are correct
@nomnamecpp3 жыл бұрын
Don't call it calibration It's calibration check
@InstrumentCalibration3 жыл бұрын
Yes... you are right. It is Verification Test
@janetawuah25353 жыл бұрын
Can you please explain the difference? I'm a beginner.