Mechanical Micrometer Calibration and Measurement Accuracy - Metrology Training Lab

  Рет қаралды 25,257

Mitutoyo America Corporation

Mitutoyo America Corporation

5 жыл бұрын

In episode 012 of the Metrology Training Lab, we'll explain the accuracy and micrometer calibration of mechanical outside micrometers. You may be surprised to find out the accuracy of a mechanical micrometer with a graduation of 0.001” is one-tenth of that.
To calibrate micrometers, and trying to decide if they are in or out of tolerance, there is another very powerful trick you need to know. *This works for Mitutoyo micrometers. The width of the scale lines on the sleeve and thimble are specifically designed to help you make higher accuracy measurements without having to read between the lines. Watch the video to find out how.
#metrology #calibration #micrometer #mitutoyo

Пікірлер: 15
@jpf1950
@jpf1950 18 күн бұрын
Thank you Jim. That was so instructive.
@sonumandlik5185
@sonumandlik5185 5 жыл бұрын
Plz make video on bore dial gauge.
@gazzzada
@gazzzada Жыл бұрын
Measurement accuracy is not a function of resolution, no matter between lines or digital. There are resolution requirements in accuracy classification but repeatability, reproducibility and bias are major contributors in any measurements including linear ones.
@jillbluerei4806
@jillbluerei4806 2 жыл бұрын
Can you use feeler guages to verify the accuracy of a ball micrometer?
@brucewilliams6292
@brucewilliams6292 2 жыл бұрын
This was very informative! Thank you.
@MitutoyoAmerica
@MitutoyoAmerica 2 жыл бұрын
Glad it was helpful!
@terrywilder9
@terrywilder9 Жыл бұрын
@@MitutoyoAmerica Here's a trick worth noting to estimate between graduation marks to get an extra order of precision: The human eye will recognize small whole fractions between these marks, the recognition of which decreases as the denominator increases. For example if you have an indication that appears to divide adjacent graduations/intervals into three equal parts this recognition will pop right out, so use/record 0.33 or 0.67, not 0.3, or 0.7, respectively; or you could have an indication that divides adjacent graduations/intervals into four equal parts use/record 0.25, 0.50 or 0.75 not 0.2 (or 0.3), 0.5 or 0.7 or 8. Then round as necessary.
@TheOne8017
@TheOne8017 Жыл бұрын
I didn't need a history lesson on micrometers. I came to this video for the exact thing that it was titled, calibration. This video showed nothing about how to calibrate a micrometer.
@elanjacobs1
@elanjacobs1 5 жыл бұрын
This is why I prefer mechanical mics over digital. If the lines are fine enough, you can even split the vernier scale to read half microns/tenths.
@havefunbesafe
@havefunbesafe 2 жыл бұрын
Question for anybody; how does a surface plate work in the world of calibration of tooling like micrometers, dial indicators, etc…I’m new to this stuff and wondering why a flat surface plate is needed for anything?!?? Pardon my ignorance on this subject. Thanks.
@MitutoyoAmerica
@MitutoyoAmerica 2 жыл бұрын
For more information on Surface Plates, please check out this video kzbin.info/www/bejne/ooi3dHSCbpl1adk
@samtzam3774
@samtzam3774 4 жыл бұрын
If someone take a measurement by guessing (aka higher than the resolution). He/She have to write it down on the recording. So if the meas is 2.2505" and the resolution of the instrument is .001" the record must be written as 2.2505"±.001". Just to give the information of the uncertainty that the record of the measurement has.
@dr.jimsalsbury695
@dr.jimsalsbury695 4 жыл бұрын
In my opinion, the more important point is to understand the difference between resolution and the scale interval (or graduation). As shown in this video, a micrometer with 0.001" graduations has a resolution 10 times smaller than that (maybe more).
@MF175mp
@MF175mp 3 жыл бұрын
@@dr.jimsalsbury695 great to see the man himself replying to comments.
@petermoses5095
@petermoses5095 5 ай бұрын
i would not waste time with electronic mikes . i have had so many errors with mikes and verniers .
Calibrate Mitutoyo Micrometer - High Accuracy MDH Outside Micrometer
7:01
Mitutoyo America Corporation
Рет қаралды 14 М.
Caliper Calibration - How to Calibrate a Caliper
17:50
Mitutoyo America Corporation
Рет қаралды 173 М.
Mama vs Son vs Daddy 😭🤣
00:13
DADDYSON SHOW
Рет қаралды 43 МЛН
КОМПОТ В СОЛО
00:16
⚡️КАН АНДРЕЙ⚡️
Рет қаралды 31 МЛН
ЧУТЬ НЕ УТОНУЛ #shorts
00:27
Паша Осадчий
Рет қаралды 10 МЛН
Understanding Metrology Measurement Units - Inch & Metric
11:54
Mitutoyo America Corporation
Рет қаралды 84 М.
Outside Micrometer Calibration - How to Calibrate - Mitutoyo
16:12
Mitutoyo America Corporation
Рет қаралды 116 М.
Metrology Quality Rules Tur-Tar
10:13
Mitutoyo America Corporation
Рет қаралды 23 М.
Micrometer Basics:  Use, Care and Calibration
20:30
Keith Rucker - VintageMachinery.org
Рет қаралды 126 М.
Gage Block Introduction - How To Use and Calibrate Gage Blocks
17:00
Mitutoyo America Corporation
Рет қаралды 88 М.
Adam Savage's Most Precise Measurement Tool!
35:21
Adam Savage’s Tested
Рет қаралды 275 М.
Influence of Temperature on Measurement - Metrology Training Lab
11:10
Mitutoyo America Corporation
Рет қаралды 18 М.
GD&T: Inspecting Position Tolerance with Bonus Tolerance Calculation
17:22
Job Shop Measuring & Metrology Tips with Mitutoyo!
25:02
NYC CNC
Рет қаралды 166 М.
АЙФОН 20 С ФУНКЦИЕЙ ВИДЕНИЯ ОГНЯ
0:59
КиноХост
Рет қаралды 1,2 МЛН
#samsung #retrophone #nostalgia #x100
0:14
mobijunk
Рет қаралды 12 МЛН
Kumanda İle Bilgisayarı Yönetmek #shorts
0:29
Osman Kabadayı
Рет қаралды 2,2 МЛН
S24 Ultra and IPhone 14 Pro Max telephoto shooting comparison #shorts
0:15
Photographer Army
Рет қаралды 10 МЛН