"Thinks about it's life, decides if it wants to work" That's literally every device in a lab, especially whenever you need that device specifically. xD
@vinicius_stelet2 жыл бұрын
Never though it was so easy to check calibration from our pipettes using only water and a balance. Thanks for sharing.
@wahajaijaz7690 Жыл бұрын
what's the acceptable range of deviation in ml of pipette and grams shown in balance.
@lacyrmoss-phdmls11 ай бұрын
It depends on the manufacturer and the volume of the pipette. Each pipette will come with an insert that states the acceptable volume for the pipette at the high end and low end of the pipette. You can then convert that volume to grams based on the density (1.000 g/mL) and temperature of the water you are using.
@raviuniqueVlogs4 ай бұрын
Are you changing the boats after?
@beautynaturefaza80526 ай бұрын
Hi Lacy, may i know how do you calculate the density of water during testing? Any simple calculation? Kindly help.