tinySA calibrating the high input

  Рет қаралды 19,253

Erik Kaashoek

Erik Kaashoek

3 жыл бұрын

A short video describing how to do the level calibration of the high input using the calibrated low input
For more information go to tinysa.org

Пікірлер: 29
@dschmider
@dschmider 3 жыл бұрын
Hi, I just had mine arrive, and I have to say I'm really happy with it. Would I be correct in saying that if I bought second TinySA, I could use the Siggen function with the method above? So far, as I am measuring gfsk transmitters at 869.75MHz with a 40dB attenuator its reading pretty good, but I'd like to be a little more certain. Cheers!
@ErikKaashoek
@ErikKaashoek 3 жыл бұрын
Yes, you can use the low output from one tinySA to calibrate the high input from another tinySA. Keep in mind the high input is factory calibrated and probably correct within 3dB
@W4BTK
@W4BTK 2 жыл бұрын
Erik thanks for the video. One thing I did also was, using a service monitor such as the HP 8920A, to generate a signal at -30dBm, I calibrated both low and high inputs. Is this a valid process after performing the Initial calibration process highlighted on the tinysa website?
@ErikKaashoek
@ErikKaashoek 2 жыл бұрын
Yes, very good!
@zeljkoadzic1692
@zeljkoadzic1692 2 жыл бұрын
new firmware not have actual power
@aciscarcab
@aciscarcab 3 жыл бұрын
What do you think about using the 300MHz armonic (from the 30MHz Cal. output build in) to calibrate? When you activate the 30MHz Cal. output, and you put the SMA bridge, the same signal is present in both High and LOW inputs.
@ErikKaashoek
@ErikKaashoek 3 жыл бұрын
That was what I was hoping for but due to internal reflections and differences in drive power of the 30MHz clock output the 300MHz harmonic has an unpredictable difference in power level at the high input versus the low input so the factory calibration has more accuracy and I decided not to use the 10th harmonic. (the same is true for the other harmonics reaching the high/low overlap range)
@aciscarcab
@aciscarcab 3 жыл бұрын
@@ErikKaashoek Thank you. Yes, the factory calibration is quite accurate.. The deviation from factory calibration was only 1 dB. After a right calibration with an external source , I see that (in my TinySA) the measured difference of these harmonics in the overlap range (240, 270 ,300, 330 MHz) is between 5.5 and 6 dB higher in the high input. If that were the same for all the TinySA´s... can create a rule to calibrate.
@ErikKaashoek
@ErikKaashoek 3 жыл бұрын
@@aciscarcab it's not the same for all
@ErikKaashoek
@ErikKaashoek 3 жыл бұрын
@@aciscarcab Its not the same in all tinySA
@chrisgroenewegen3756
@chrisgroenewegen3756 3 жыл бұрын
Hi Erik, I'm owning a SDG2042X from Siglent. Which signal generator is able to provide a known signal of e.g. 250Mhz? Those would be I assume quite expensive? Looking for ways to calibrate the high input too.
@ErikKaashoek
@ErikKaashoek 3 жыл бұрын
In general the high input is without calibration accurate within 2dB so no real need to calibrate. You can however calibrate using any stable signal between 240MHz and 350MHz. Even a DBV-C or ATSC cable TV signal will work. Using your SDG2042X you can program a signal with a LOT of harmonics and use one of the harmonics to do the calibration.
@chrisgroenewegen3756
@chrisgroenewegen3756 3 жыл бұрын
@@ErikKaashoek Ah Erik, I now remember you made a comment on taking a std TV signal from ATSC indeed. Thanks for the tip regarding the SDG2042X appreciate it. Still many things to learn and no time. ;-)
@BenHydeSPARKScience
@BenHydeSPARKScience Жыл бұрын
I'm totally lost... At the 1:00 mark, you say that the "signal generator is hooked up..." How do you do this?
@ErikKaashoek
@ErikKaashoek Жыл бұрын
Any signal from any device supplying a stable signal between 240MHz and 350MHz can be used. Can even be a Cable TV signal.
@Ibelieve218
@Ibelieve218 3 жыл бұрын
I have no signal in the required band for teh high.I checked on signal generators there not cheap and i really dont want to drive around looking for a signal seems like thats the only draw back
@ErikKaashoek
@ErikKaashoek 3 жыл бұрын
Yes, but the high input is (as explained in the wiki) for free so it has its limitations. The measurement error is anyway less then 2dB when uncalibrated
@stevencraig
@stevencraig 2 жыл бұрын
Hi Erik, I no longer see the "Actual Power" setting on the expert config menu. Has it been removed or renamed? My firm we are is v1.3-265
@ErikKaashoek
@ErikKaashoek 2 жыл бұрын
The actual power is moved to a submenu
@stevencraig
@stevencraig 2 жыл бұрын
@@ErikKaashoek I never thought to look in a submenu. I've found it now, thanks. Is the manual on your website editable like a wiki, (i.e. open source)? It still reflects the old menu structure and I can look into updating it.
@ErikKaashoek
@ErikKaashoek 2 жыл бұрын
@@stevencraig Yes, the wiki is user editable. Contact me via email and I will provide you with the details
@Ibelieve218
@Ibelieve218 Жыл бұрын
Newer one does not have power I have level in or output??? So how do you calibrate this one??
@ErikKaashoek
@ErikKaashoek Жыл бұрын
I do not understand your question
@Ibelieve218
@Ibelieve218 Жыл бұрын
@@ErikKaashoek mine I just bought recently. If you go into config, more I have Level correction don't see power level setting. There are no power level corrections
@Ibelieve218
@Ibelieve218 Жыл бұрын
in level correction there's setting for input level and output level
@ErikKaashoek
@ErikKaashoek Жыл бұрын
Please join this group for support groups.io/g/tinysa/messages
@Ibelieve218
@Ibelieve218 3 жыл бұрын
I don't have a signal generator you said I can use low to get the high done
@ErikKaashoek
@ErikKaashoek 3 жыл бұрын
You can use any stable signal that is within the frequency range of both the low and high input. It could even be a cable TV signal. You can NOT use the harmonics of the calibration output.
@Ibelieve218
@Ibelieve218 3 жыл бұрын
@@ErikKaashoek thank you I found one set it for center then went to high and followed your direction thank you
tinySA Mistakes
11:56
Erik Kaashoek
Рет қаралды 31 М.
tinySA Level Menu
11:37
Erik Kaashoek
Рет қаралды 5 М.
Became invisible for one day!  #funny #wednesday #memes
00:25
Watch Me
Рет қаралды 59 МЛН
УГАДАЙ ГДЕ ПРАВИЛЬНЫЙ ЦВЕТ?😱
00:14
МЯТНАЯ ФАНТА
Рет қаралды 3 МЛН
I Can't Believe We Did This...
00:38
Stokes Twins
Рет қаралды 124 МЛН
#558b TinySA High Port Calibration (It's not easy)
15:18
IMSAI Guy
Рет қаралды 6 М.
tinySA doing AM and FM generation and analysis
11:36
Erik Kaashoek
Рет қаралды 10 М.
The spectrum analyzer explained
12:25
RCModelReviews
Рет қаралды 52 М.
TinySA ULTRA_UK
16:38
Kurt Poulsen
Рет қаралды 18 М.
#597 Spectrum Analyzer Basics (part 1)
14:27
IMSAI Guy
Рет қаралды 53 М.
tinySA Ultra Spectrum Analyzer Review/Experiments/Teardown
29:05
TinySA - EP3 - Ajuste do High Input
4:26
W4BTK
Рет қаралды 1,2 М.
Intro to Spectrum Analyzers - What does a TinySA do?
19:07
Temporarily Offline Ham Radio
Рет қаралды 27 М.
Обзор анализатора спектра TyniSA