Press "set" to enter the setting interface, and press "down" to enter the parameter setting.
Press and hold "up" and "exht" at the same time, and then enter the engineer password in order: 556677 (press "exht" after each digit to switch to the next digit).
Then press "set", and then press "down" to select "model selection". The last interface you enter is the engineer system, which only has a Chinese interface.
Don't worry, keep pressing "down" until you jump to "Smart cleaning: yes", then press "set", when "yes" flashes, press "up" again and it will become "no" (off), and then press "set" to confirm the setting.
Finally, press "data" continuously to return to the main interface of autoclave.