Analog ammeter calibration _Failed
11:50
Adaptor performance Testing
10:36
3 ай бұрын
Metravi LCR 460 calibration
2:25
4 ай бұрын
Panel Resistance meter demo
7:22
5 ай бұрын
Ammeter - How to kill it!
9:01
9 ай бұрын
Amps measurement using Fluke 106
3:45
Earth megger calibration on site
20:27
Selec voltmeter MV15 usage
8:01
Жыл бұрын
Multimeter auto calibration.
3:44
Пікірлер
@sharirk4529
@sharirk4529 4 күн бұрын
I need a meter to measure less than 200ma . Can you show like dc ma
@sharirk4529
@sharirk4529 4 күн бұрын
Can I check DCma.means less than 200
@30mAkills
@30mAkills 3 күн бұрын
You have to buy AC and DC leakage current meter. Read the specs and ensure that the model has DC feature before buying.
@Dog-whisperer7494
@Dog-whisperer7494 9 күн бұрын
I been using the exact same multimeter for for y to thirty eight years and I have never had any problems and it has never let me down and I even test three phase supply with it as well as 240v single phase supply.
@30mAkills
@30mAkills 8 күн бұрын
Agreed. But now components quality are compromised, especially caps.
@SVsaelect
@SVsaelect 21 күн бұрын
Thank u very much sir
@SVsaelect
@SVsaelect 21 күн бұрын
Sir good evening ❤
@30mAkills
@30mAkills 21 күн бұрын
Speech error @7:25 to 7.35 - 4.6Amps NOT 5.6 Each division value is 0.1Amps NOT 0.2Amps
@SVsaelect
@SVsaelect 29 күн бұрын
Sir neenga palayamadhri Tamila video podunga sir plz
@seshivamenterprises111
@seshivamenterprises111 Ай бұрын
Charging Transformer kitne volt ka laga hota hai
@seshivamenterprises111
@seshivamenterprises111 Ай бұрын
Waco earth tester mein charging led on ni ho rha..
@abdulazeez4930
@abdulazeez4930 Ай бұрын
How to test smps sir
@30mAkills
@30mAkills Ай бұрын
By using DC load bank and measure voltage.
@SVsaelect
@SVsaelect Ай бұрын
Purchase link sir
@30mAkills
@30mAkills Ай бұрын
toolworld.in
@30mAkills
@30mAkills Ай бұрын
Safety warning: Always discharge capacitors before testing and after testing using suitable resistors (another subject). A charged capacitor is like a bomb. Do not short capacitors.
@SVsaelect
@SVsaelect Ай бұрын
Useful sir
@SVsaelect
@SVsaelect Ай бұрын
Very useful sir
@SVsaelect
@SVsaelect Ай бұрын
Sir best capacitor check meter best model
@30mAkills
@30mAkills Ай бұрын
HTC CM 1500/1501
@SVsaelect
@SVsaelect Ай бұрын
Very useful sir
@morrisignacio9351
@morrisignacio9351 Ай бұрын
If you can use it for a year then your lucky😂😂😂
@murugans-el8np
@murugans-el8np Ай бұрын
நீங்கள் சொல்வது சரிதான்.. ஆனால் எந்தக் குறையும் இல்லாமல் 14 ஆண்டுகளாக உபயோகிக்கிறேன்.. 2 அல்லது 3 வருடங்களுக்கு ஒரு முறை புதிது வாங்கி விடனும்...ரூ 200 தானே.
@30mAkills
@30mAkills Ай бұрын
But this is new purchased from Robu.in Sealed packing.
@hansmoser162
@hansmoser162 2 ай бұрын
And what about the decal? For proper mounting it should be removed.
@30mAkills
@30mAkills 3 ай бұрын
In this case suppose the voltage drops to 8.5 V or below @ 1 amps draw, means the adaptor cannot deliver full power and its useless, cheating by the manu. You can use automotive lamps as load instead of resistors.
@SVsaelect
@SVsaelect 3 ай бұрын
Sir iwant ac capacitor check meter best and cheap which one you r choice
@30mAkills
@30mAkills 3 ай бұрын
HTC CM 1500
@SVsaelect
@SVsaelect 3 ай бұрын
Kishore sir very useful
@SVsaelect
@SVsaelect 3 ай бұрын
Sorry sir romba nalla video podava illa ❤❤❤
@SVsaelect
@SVsaelect 3 ай бұрын
Why sir long video
@30mAkills
@30mAkills 3 ай бұрын
How many adopters you have killed?
@shivaprasanth952
@shivaprasanth952 3 ай бұрын
Sir is this good. Can i purchase and what is the difference between htc emf523 and this
@30mAkills
@30mAkills 3 ай бұрын
For this price we don't get calibration certificate.
@rounakr94
@rounakr94 3 ай бұрын
Your N-E voltage seems a bit high like mine. Mine is around 6V in the evening and 1V in the morning. Is it okay?
@30mAkills
@30mAkills 3 ай бұрын
OK Try wetting the earth pit.
@mohanbabu3677
@mohanbabu3677 4 ай бұрын
You put primary side high current load like equal to 100amp .if 100/5 or 100/1 amps ct means you put more current in primary. 200w load not affected
@30mAkills
@30mAkills 4 ай бұрын
Note that there is no raise in voltage. If you have ref source please post the link. We believe sun raises in the east and sets in the west!
@bijuarjun
@bijuarjun 4 ай бұрын
Thank u Sir...❣
@30mAkills
@30mAkills 4 ай бұрын
This is an accurate, tough and waterproof (IP 67) LCR meter. I recommend.
@allzmartnetworkprivatelimi7472
@allzmartnetworkprivatelimi7472 4 ай бұрын
hi can you share your number
@bannisawariya8821
@bannisawariya8821 5 ай бұрын
Well done sir 🎉🎉
@argkgyan2244
@argkgyan2244 5 ай бұрын
Tum ho india me video bana rahe ho English me .hindi me banao
@evmohanan2658
@evmohanan2658 5 ай бұрын
Valuable information ❤ well 😊
@shaikmeharaj5011
@shaikmeharaj5011 5 ай бұрын
Nice sir thank u
@30mAkills
@30mAkills 6 ай бұрын
I like the dual colour display feature. If the tested circuit parameters are OK, the display is green. If it senses faults like reverse polarity, missing earth then the display turns red. This ensures that we won't miss faulty circuits.
@kharna_feed
@kharna_feed 6 ай бұрын
How much time will press the test button for reading
@30mAkills
@30mAkills 6 ай бұрын
Until reading is stable.
@bobyantony2937
@bobyantony2937 6 ай бұрын
Gud one sir ❤
@SVsaelect
@SVsaelect 7 ай бұрын
How many second rotate hadle sir
@30mAkills
@30mAkills 7 ай бұрын
Refer user manual. Normally 160rpm
@SVsaelect
@SVsaelect 7 ай бұрын
Sir which side input or out put...for rccb
@30mAkills
@30mAkills 7 ай бұрын
According to labelled diagram
@SVsaelect
@SVsaelect 7 ай бұрын
Netural earth link aanulum light eariumeay sir...?
@SVsaelect
@SVsaelect 7 ай бұрын
Sir iam felix from ooty...rcd available a sir please rate and details and rccb and rcd different sir
@30mAkills
@30mAkills 7 ай бұрын
RCCB and RCD are same. About rate enquire in electrical shops.
@Blathurkaran
@Blathurkaran 7 ай бұрын
Sir what happened when we set 10 ohms
@30mAkills
@30mAkills 7 ай бұрын
Like multi meter display will show OL if the resistance value exceeds 10 ohms. If you mean "get" instead of "set" then 10 ohms is acceptable. Sorry for late response.
@Blathurkaran
@Blathurkaran 7 ай бұрын
@@30mAkills I have get reading 01.8 is it good
@30mAkills
@30mAkills 7 ай бұрын
@@Blathurkaran what's the resistance value of lead wires? Also study FOP method from AEMC/Fluke/Megger knowledge base which are freely available at their web sites. FOP graph will be useful. Hope this helps.
@30mAkills
@30mAkills 9 ай бұрын
IR Testers or Meggers are widow-makers unless you follow the below safety precautions: 1. Always wear gloves. 2. Check the probes and leads visually for damages like cracks, frayed necks and exposed wires. 3. Ensure that the device or circuit under test is not energized. 4. After testing, discharge the residual voltage retained by the device (cables, motors and transformers due to capacitive effect) simply by grounding the device terminals for a minute.
@akhanani1918
@akhanani1918 4 ай бұрын
Kishore ji, ur contact no. pliz need to talk to you about power plant equipments.
@SVsaelect
@SVsaelect 9 ай бұрын
Meter price sir
@30mAkills
@30mAkills 9 ай бұрын
Check in amazon
@30mAkills
@30mAkills 9 ай бұрын
You can also destroy a multimeter by trying to measure volts when the probes are plugged into the amps port!
@statuztamila353
@statuztamila353 9 ай бұрын
Enaku unga videos paathu learn pannikanum nu aasai iruku but language problem ah iruku🤦🏻‍♂️😢
@30mAkills
@30mAkills 9 ай бұрын
DUT: HTC CL 2054 Leakage current Clamp Tester. Purpose of video: Calibration of mA range. Verdict: Readings are within tolerance limits. Hence Recommended. Uses: Leakage current testers are commonly used for measuring leakage currents (differential) which are present in mA range in any electrical circuit or equipment, which cannot be sensed by normal clamp meters. At higher ranges say 1 amps and above the leakage tester can be used as normal clamp meter. Usage method for measuring leakage current: Clamp all current carrying and neutral conductor (if present) together between the jaws or only the earth wire of the circuit. Precautions: 1) The wires should be approximately at the centre of the jaws. 2) The jaws should be closed firmly without any gap. 3) Wear gloves for safety. Causes of Leakage current: Insulation failure, capacitance, filters, improper earthing, are some of the common reasons. Effects of leakage current: 1) Can cause nuisance tripping of RCCB's. 2) Fire. 3) Electrocution during monsoon. Audio Correction @ 6:39 = RCCB and not MCB
@30mAkills
@30mAkills 9 ай бұрын
DUT: HTC CL 2054 Leakage current Clamp Tester. Purpose of video: Calibration of mA range. Verdict: Readings are within tolerance limits. Hence Recommended. Uses: Leakage current testers are commonly used for measuring leakage currents (differential) which are present in mA range in any electrical circuit or equipment, which cannot be sensed by normal clamp meters. At higher ranges say 1 amps and above the leakage tester can be used as normal clamp meter. Usage method for measuring leakage current: Clamp all current carrying and neutral conductor (if present) together between the jaws or only the earth wire of the circuit. Precautions: 1) The wires should be approximately at the centre of the jaws. 2) The jaws should be closed properly without any gap. Causes of Leakage current: Insulation failure, capacitance, filters, improper earthing, are some of the common reasons. Effects of leakage current: 1) Can cause nuisance tripping of RCCB's. 2) Fire. 3) Electrocution during monsoon. Audio Correction @ 6:39 = RCCB and not MCB
@30mAkills
@30mAkills 9 ай бұрын
Don't do this if you don't have the guts😮 So we should not measure volts when the probes are plugged into the amps port. Great lesson.
@KamalSingh-qy2dd
@KamalSingh-qy2dd 9 ай бұрын
What is E1 and P1 and E2 and P2