Yeah they should be nearly the same - some of the boards are a bit different looking - depending on when the amx4+ was built but overall pretty much the same. Of course it’s always good to check the documentation for the model your using
@engr.israrahmad31946 ай бұрын
Thanks so much How to calibrate drive handle
@HTMWorkshop6 ай бұрын
GE has a documentation portal - you'll need to get the service manual for the AM4+ and use that to calibrate the drive handle - its not to hard, its a pretty quick calibration. If you know how to get to the service mode of the machine it will sort of walk you through it on the prompts.
@muhammadsajjadaslam64643 ай бұрын
When i power up machine i get error XRAY DISABLED I JUST REPLACE NEW BATTERY
@muhammadsajjadaslam64643 ай бұрын
@if possible may i have your contact??
@Realgrowinstant2 ай бұрын
My Amx-4 shows (disabled xray) & ( dives only) please help me
@HTMWorkshop2 ай бұрын
You have to do all the calibrations to restore functionality - including taps. Once you start the generator calibration you have to do all the others. Each calibration builds off the previous one so make sure you do them in the correct order
@rolandocanizavillanueva11102 жыл бұрын
when I try to do the calibration mAs I get Xrax Word limit. And it does not allow me to continue with the calibration kvp
@HTMWorkshop2 жыл бұрын
I have not had to deal with that one - I know there are some ideas on Medwrench for troubleshooting that issue but not sure its the same. Sounds like an odd one.
@HTMWorkshop2 жыл бұрын
If you figure it out please let me know
@samherzog-yv6mt Жыл бұрын
I had this same problem and found out if you adjust the kvp values by more than three each time the old memory chips hit a limit and kick you out of calibration. So only adjust by three (full numbers) max and it will take more X-rays but will complete the KVP cal in the end!
@reubenmurey88763 жыл бұрын
How can I solve the problem of blunt image on my Ge Amx4. The image is bad
@HTMWorkshop3 жыл бұрын
Not sure on that one - have you checked any of the forums such as medwrench. You could also try asking r/BMET to see if they have any suggestions on reddit.
@equaye69822 жыл бұрын
can i calibrate without the raysafe? I dont have one.
@HTMWorkshop2 жыл бұрын
I could be wrong but I believe there is a way to do it with an oscope and a voltage divider but it’s been a while - I think it’s in the manual if my memory serves
@equaye69822 жыл бұрын
@@HTMWorkshop Thanks so much for your prompt reply. I have got the calibration manual, i will check.
@muhammadsajjadaslam64643 ай бұрын
@ 8:32 HTM WORKSHOP WHEN WE ON THE MACHIEN TEST 1 OK TEST 2 ON TEST 3 OK AFTER THAT XRAY DISABLED APPEARS
@Andres-eb9gn3 жыл бұрын
When I try to calibrate Kvp, I reach 120 kvp but the machine turn off and restart the system for the begining
@HTMWorkshop3 жыл бұрын
Interesting - I have not dealt with that issue before. Is the battery fully charged?
@HTMWorkshop3 жыл бұрын
Also looks like someone else has had the same issue here but no clear resolutions - www.medwrench.com/thread/149877/amx-4-issue
@qandeelnisar56383 жыл бұрын
Aoa Sir From Askari Diagnostic lab Narowal Sir yesterday machine was charged on 1/4 charging.when charging was completed breakers was tripped and Now machine is giving the option (Test: 3 Failed)
@HTMWorkshop3 жыл бұрын
Not sure - but the service manual does have troubleshooting guide in the back that you could try based off the thrown error code. Wish I could help you - let me know if you find a solution!
@joelohamagor94624 жыл бұрын
PLEASE HOW CAN I SOLVE ERROR 45B ON AMX4 ..PLEASE
@HTMWorkshop4 жыл бұрын
Have your tried any of these fixes yet? www.medwrench.com/thread/120576/amx-4-intermittant-45b-error www.dotmed.com/virtual-trade-show/category/Service-Technicians/Discussions/Forums/Amx-4-Plus-Error-Code-45/8286
@luisrosalesnft4 жыл бұрын
If you have the error on the calibration of the generator revise the condition of your fil/kvp board first
@qandeelnisar56383 жыл бұрын
How can we solve out (test failed 2....then test failed 3] error