A demonstration of how to use a Fluke model 744 Documenting Process Calibrator to auto-calibrate an I/P converter (4-20 mA in, 3-15 PSI out).
Пікірлер: 6
@MattSyTy12 жыл бұрын
That's quite the piece of equipment you've got there. I'd have to really make a good case to requisition one of these bad boys.
@lloydrobles12 жыл бұрын
pretty cool thxs, again man.for this short videos.
@kevingawada961410 жыл бұрын
Thanks for the knowledge you share on these short videos. I have a question on this particular one. What instrument is pumping the pressure up and how is it being controlled by the 744. I can see that you are sourcing the 4-20ma and i can also see the connection to the pressure module but trying to figure how the pump is being controlled Cheers Kev
@bdom31065 жыл бұрын
There is no pump. In this video he is sourcing the positioner I/P with the Fluke 744. The I/P (transducer) is the pressure source. He is using a Fluke pressure module which is tied into the same line as the I/P output for pressure feedback to the 744
@strider001115 ай бұрын
The input if factory air to the I/p, the output is regulated by the 744..
@iwecotechingenieriayservic52748 жыл бұрын
thanks for video. I have a question on this particular one. instrument which I use to calibrate the Fluke 744? greeting