I purchased the RC5+ for continuous temperature monitoring for storage of COVID vaccine, as required by the CDC and the NYCDOH. I am quite satisfied with the device, found its set-up process relatively easy, accompanied by this video and am able to upload data for submission to my health department.
@ElitechTechnologyInc3 жыл бұрын
Hi, we really appreciate your feedback. Any problems, please contact us freely.
@windkl2 жыл бұрын
Extraordinary device with exceptional accuracy in measurements!
@SonicOrbStudios11 ай бұрын
Love these for diagnostic in refrigerators at my apartment complex
@vincebush2 жыл бұрын
I could not find the Thermal Time Constant info on any data sheet so I took a chance and purchased one. I measured a whopping 8 minute Time Constant. It take 40 minutes to reach a new temperature. FAR too slow for what I need so I returned it.
@unionse7en2 жыл бұрын
any way too download csv data to a mobile phone in the field, rather than a laptop?
@starhopper457 Жыл бұрын
I now have one and so far love it. A question though - can it store more than one program and then I pick which one to run? I may need it for an astronomical event where I may not have immediate access to a computer to adjust things, so having multiple templates would be nice.
@anaisabelazevedogregorio9538Ай бұрын
good morning Is it possible for the data logger to connect to the USB port and immediately open a window with the report, without the need for this computer to have Elitech software? At my work, the computers do not allow you to install software outside the service. I wanted to use this data logger to monitor a vaccine refrigerator. thanks
@Neinei09132 жыл бұрын
hello, device shows 6/21/2021, how to change to current date and time? thanks
@Carla-hq6yp Жыл бұрын
Does the RC-5+ have a Bluetooth option to retrieve temperature log?
@cwilliams94003 жыл бұрын
Should I leave it in the fridge and it will regularly monitor the temp?
@ElitechTechnologyInc3 жыл бұрын
Hi, that's right! The logging interval can be set in our free software according to your preference.
@MohamedAhmed-nh6ef Жыл бұрын
Is this data logger reusable or it will log for certain period then it life end like 60 days or 90 days logger ?
@Carla-hq6yp Жыл бұрын
Do you have a Elitech temperature logger/device with a Bluetooth option?
@PaulaAndreaAriasArellano Жыл бұрын
Hello, I need help, the pause and play symbols appear on the LCD screen of the RC5+ at the same time. How can it be removed to start recording data???
@ronkloiber2 жыл бұрын
Hello, I have the RC-5+ I am able to start recording after I have set the parameters in software, but if I stop the recording, I am not able to start it again without saving parameters to it again. Is this normal? if not, how can I start and stop multiple times? Does it append to an existing log, or start a new file?
@auniesakinahh30303 жыл бұрын
sorry hi , how to set up the date .
@ElitechTechnologyInc3 жыл бұрын
Hi! To setup the data and time, please connect the device to your computer, open the software ElitechLog. Go to Parameter, and select the correct time zone and click Save Parameter in the end. If the time zone setting field is greyed out, please click Save Parameter directly, it will sync to your computer time automatically.
@windkl2 жыл бұрын
Could you provide info about the response time of temperature measurements?
@ElitechTechnologyInc2 жыл бұрын
The display screen refreshes the temperature once in about 3-5 seconds.
@ronaldreuling610311 ай бұрын
40 minutes
@DIVA17MARIAN2 жыл бұрын
Hello I have a question about the Calibration (temperature) parameter, how to use it?
@ElitechTechnologyInc2 жыл бұрын
An offset (not slope) calibration is possible using the RC-5 software. With the device connected to a USB port (and the software permitted to "make changes to this computer" despite the "unknown publisher" or missing certificate): Click "Parameter Set" (in toolbar or Settings menu), then enter the required delta (negative to decrease the reading) as "TemperatureCalibration" near the bottom of the second column, then click button "Save Parameter", confirm and Exit. Note that although the Up/Down arrow buttons change the offset by a full degree, you can edit to the displayed resolution of 0.1°. For any further questions, you may have a look Elitech FAQ part.
@coolmonkey6192 жыл бұрын
Hi can I use this as a stand alone room sensor.
@ElitechTechnologyInc2 жыл бұрын
Hi, sure, they canwork as room temperature monitors.
@meme-mm2wp2 жыл бұрын
Hi, is the device can be calibrated?
@ElitechTechnologyInc2 жыл бұрын
Yes, you could send the unit to your local calibration center for the re-calibration service.
@caslamdq3 жыл бұрын
How do I have to set the parameters to get an hour-by-hour daily record during a month?
@ElitechTechnologyInc3 жыл бұрын
Hi! Are you trying to set the logging interval to 1 hour? If so, please connect the device to your computer, open the software ElitechLog, go to Parameter, find Logging Interval and set it to 1 hour 0 min 0 sec. Click Save Parameter in the end. Then you device will be ready to start a new logging with 1-hour recording interval.
@caslamdq3 жыл бұрын
@@ElitechTechnologyInc Hi, thanks for your answer!. Yes, I have set the logging interval to 1 hour. However, it recorded during 6 days and wrote one register. Thus, I have just one temp Max and Min. What I need is a daily record with its Max and Min records. Is this possible?.
@caslamdq3 жыл бұрын
@@ElitechTechnologyInc Hi, thanks for your answer. What I need is to get a daily record with its Max And Min Temp for at least 1 month.
@cburshtein2 жыл бұрын
I just got mine but it's showing NC and the play+stop icon are on and off constantly. It doesn't start when the play button is pressed for 5 seconds. Where can I get assistance?
@ElitechTechnologyInc2 жыл бұрын
Hi Cecilia, the NC message appears with two possible reasons. One is the temperaure is out of the measurement range. Another one is the sensor issue. You may contact our support team via support@elitechus.com
@windkl2 жыл бұрын
probably it is configured through the app to "external" sensor type. You need to change the sensor to "internal" .
@cburshtein2 жыл бұрын
@@windkl It was not, it continued to go off and on randomly.
@windkl2 жыл бұрын
@@cburshtein I see. Then it's definitely defective.
@muhammadarfanulaziz15073 жыл бұрын
it can be submerge in liquid for 24 hour?
@ElitechTechnologyInc3 жыл бұрын
Hi Arfanual! No, they cannot be submerged into the liquid.
@ablosport3122 жыл бұрын
Can it measure temperature from +2 to +8°c?
@ElitechTechnologyInc2 жыл бұрын
Hi, yes, the measurement of RC-5+ and RC-5+ TE is both -30°C~+70°C.
@ablosport3122 жыл бұрын
@@ElitechTechnologyInc thank you very much..
@FB-gm6el Жыл бұрын
i wish things like "low battery!" and "recording in process" were a bit clearer/more intuitive...