This is really great! Thanks! Also, I would also like to see the Fluke unboxing videos, and hope you could shot some.
@stalkinghawk92442 жыл бұрын
Absolutely brilliant
@jacobdellinger49313 жыл бұрын
Do you happen to know what hart function is for? Ty for great video
@fridayswithbrandon3 жыл бұрын
It puts a 250 ohm resistor in series if you are in source mode and activate the HART feature.
@jacobdellinger49313 жыл бұрын
What’s the reason for needing 250 res? Sorry for all the question lol
@fridayswithbrandon3 жыл бұрын
@@jacobdellinger4931 if you have HART smart sensors they will need a 250 ohm resistor to work.
@sandipdamodar87863 жыл бұрын
Hi... Need solution for Fluke 773 It's mA range has changed to 24mA, instead of earlier 20mA range.At 100 percent current is 24mA. I want it at 20mA.
@fridayswithbrandon3 жыл бұрын
I'm not sure about that one off the top of my head. But I would recommend calling Fluke's tech support at 800-44-FLUKE. They are open from 9:30 EST to 5:30 EST
@sandipdamodar87863 жыл бұрын
@@fridayswithbrandon Thnks....will check for solution....
@HuskyNoKokyuu2 ай бұрын
"Source is used if you don't have a power source, but if you don't have a power source then you would use simulate" Uhh, what?
@fridayswithbrandon2 ай бұрын
@@HuskyNoKokyuu if you do have an external power source outside of the meter then you would use simulate.