Why is kinetic energy = electric potential in this concept ?
@mazdaksheytunak6939 Жыл бұрын
Great Video! Love your channel! I have some minor corrections to your video: 1) TOF: detected m/z value is dependent on BOTH the molecular mass ("size") and the CHARGE of the analyte. it is not only dependent on the "size" as you stated. A peptide with a mass of 1 kDa, for instance, can be detected at ~ 1kDa m/z if it bears 1 charge, but also at 500 Da m/z if it bears 2 charges, and at 333 Da m/z if it bears 3 charges. 2) you should not use the term "size" when referring to the molecular mass of a molecule. These are 2 different physical properties. Thanks! :)
@LucasLearnz Жыл бұрын
Great points! Thank you so much! I've been considering to make an updated version to this video since I could make a much nicer version today so when I do I'll make sure to include your corrections!👍
@octaviorojas5522 жыл бұрын
Thanks for the video! Very clear and simple explanation
@LucasLearnz2 жыл бұрын
Thank you so much for those nice words Octavio!😇
@fazelamirvahedi9911 Жыл бұрын
Thanks a million for this fantastic educational video. I have two suggestions, I guess it would be much better if you use some pictures from real instruments which are used in laboratories, and secondly, if you produce a series of videos on using the common analyser software to teach the whole PMF process, it would be great.
@LucasLearnz Жыл бұрын
Great suggestions! I'll see what I can do!
@edreesamani22945 ай бұрын
Hello this was a great video explaining how MALDI-TOF works! I am currently using the MALDI-TOF on some peptides I'm making and I had a question. The peptides we are making has a +4 net charge but on the maldi we usually see +1 charge mass (m/z) on the spectra and not the +4 charged (m/z) mass. Can you explain this? Thank you!
@LucasLearnz5 ай бұрын
How does that show up exactly? I mean how do you know that you're seeing +1 charge rather than +4 charge?🤔 I've only analyzed things that had a +1 charge myself I'm afraid🙈😅
@edreesamani22945 ай бұрын
@@LucasLearnz Hi thanks for responding! I know that it is the +1 charge because that is the predicted mass of my peptide. My peptide has a mass of 1154 g/mol and that is the peak that I see on the maldi-tof. I am just confused because the peptide has charged amino acid residues that should be fully protonated with a net +4 charge when it is on the plate with the matrix.
@최수빈코털2 жыл бұрын
thx! it was very helpful for my presentation
@LucasLearnz2 жыл бұрын
I am happy to hear that!👍
@so.ciaaaa9 ай бұрын
once you have "digested" the proteins and splitted them into parts of different molecular weights, how do you know which parts belong to the same protein in order to have an accurate identification of it?
@LucasLearnz9 ай бұрын
This depends. Sometimes you do your best to isolate the protein as much as possible before you use MALDI-TOF. However, often you might compare the results to existing libraries of mass spectra in order to try and find a match. Does that help at all?
@KoalaKid232 жыл бұрын
Very helpful! Thanks!
@LucasLearnz2 жыл бұрын
Excellent! Glad to hear it! I also have videos about MALDI and TOF separately if you wish to know more about either one of these two techniques😇👍
@sahanar5672 Жыл бұрын
Great explanation Thankyou 🤟
@LucasLearnz Жыл бұрын
Great that I could help!
@hernandezsalmoranerik47187 ай бұрын
THX!!!! Great explanation!
@LucasLearnz7 ай бұрын
Happy I was able to help!
@michaelala47062 жыл бұрын
Thank you great video
@LucasLearnz2 жыл бұрын
Glad to hear that you found it helpful! Let me know if you have any questions or suggestions for other topics to cover!👍
@sorlos11 ай бұрын
Thank you!!
@LucasLearnz11 ай бұрын
Happy I could help!!😇👍
@noraspeiser18652 жыл бұрын
What exactly is the difference between MALDI-TOF and MALDI? Whitout measuring the TOF how does MALDI analyze the samples?
@LucasLearnz2 жыл бұрын
Excellent question! The MALDI part is ONLY responsible for ionizing the sample while the TOF is ONLY responsible for measuring the ionized sample. In other words, we can use another ionization method such as for instance electrospray ionization (ESI), which I cover in another video, instead of MALDI. In the same way, we can use another method instead of TOF to measure our sample. You can think of this as two different parts of one larger device that can be changed out depending on what samples you are measuring and so forth... First "real-life" metaphor that comes to mind would be in gaming, where you can either use a PC, Playstation or an Xbox and plug this device into a desktop screen, or some type of TV... Not the best analogy but hopefully that answers your question!
@noraspeiser18652 жыл бұрын
@@LucasLearnz perfectly explained, thank you very much!!
@LucasLearnz2 жыл бұрын
@@noraspeiser1865 Glad to hear that, thanks for commenting!!
@nailaimtiaz361711 ай бұрын
Make video on illumina sequencing
@LucasLearnz11 ай бұрын
Here you go😇 kzbin.info/www/bejne/Z5y4f3qbjJJ-gbMsi=XdXbyBtO-CfHVxZj
@amosadjei79619 ай бұрын
Nice ❤
@LucasLearnz9 ай бұрын
I'm happy you liked it!😇🙏
@lucysmith4263 Жыл бұрын
Hi I'm currently trying to analyse both positive and negative MALDITOF spectra for some lanthanide complexes I synthesised to try and determine the molecular weight and hence structure of my products, but I'm really struggling to understand how to interpret the spectra. I would really appreciate any help you can offer.
@LucasLearnz Жыл бұрын
If you mail me some of the spectra at biotechlucas@gmail.com I can take a look😇👍
@LucasLearnz Жыл бұрын
Also if you know what matrix you're using and the standard used for calibration that would help as well🙏
@shantinisureshkhanna2448 Жыл бұрын
Excellent teaching🎉Can you add a video on NMR spectroscopy?
@LucasLearnz Жыл бұрын
A short video series on different aspects of NMR will be the next thing I do! Promise! :D
@sto277910 ай бұрын
can you describe how to build one? thanks.
@LucasLearnz10 ай бұрын
Interesting question! What is the context if I may ask?😇 Do you mean theoretically or how these are built by different companies?
@AbdurrohmanZuhdi8 ай бұрын
So MALDI ionization technique is same as Chemical Ionization (CI) technique in GC mode?
@LucasLearnz8 ай бұрын
Not entirely sure what you mean. Chemical Ionization is also an ionization technique true but they are different in terms of how they are carried out and how well they preserve the sample.
@lilahmolina28452 жыл бұрын
Can you go more in-depth about MALDI?
@LucasLearnz2 жыл бұрын
Absolutely, might do it this week or the next week at the latest!