Great video sir thank you for sharing. Am happy about the English subtitles 👍👍 because I can understand the procedures
@d.i.yrepairnmaintenance9538 Жыл бұрын
sorry..I need time to make the translation..because of busy work in the field
@d.i.yrepairnmaintenance9538 Жыл бұрын
I hope that video is useful
@joelbaguma8551 Жыл бұрын
@@d.i.yrepairnmaintenance9538 yap it's so helpful to me thanks alot
@d.i.yrepairnmaintenance9538 Жыл бұрын
Thank you..for your support
@badis_dz5808 Жыл бұрын
Thanks bro 🙆
@d.i.yrepairnmaintenance9538 Жыл бұрын
You're welcome..
@icemanb5065Ай бұрын
Great video . I was dealing with a problem similar
@d.i.yrepairnmaintenance9538Ай бұрын
@@icemanb5065 thank you
@TechnoMedBME4 ай бұрын
Thank you
@d.i.yrepairnmaintenance95384 ай бұрын
@@TechnoMedBME you're welcome
@aamery7198 Жыл бұрын
Great video brother 👍
@d.i.yrepairnmaintenance9538 Жыл бұрын
Thank you brother
@jlmonteblack Жыл бұрын
Great vídeo, cómo puedo contactarlo?
@d.i.yrepairnmaintenance9538 Жыл бұрын
This my whatsapp number +62 815-4535-1558
@irzaharun46839 ай бұрын
Izin bertanya pak, kalau hasil running sampel pasien WBC, RBC, dan PLT smuanya **** itu diagnosanya apakah terdapat sumbatan di jalur probe nya ya pak? 🙏
@d.i.yrepairnmaintenance95389 ай бұрын
@IrzaHarun4683 jika disertai eror clog / re-count di akhir pembacaan sample..berarti terjadi sumbatan bisa di sebabkan oleh darah sample nya (kurang homogen)..coba masuk maintenance..zap aperture 3x.setelah itu flush aperture 3x..coba baca lagi sample nya
@andisomeng12088 ай бұрын
Mas mau tnya klau gak kluar gelembung dan sensor merahnya cuma berkedip cepat apanya y mas..mksih
@d.i.yrepairnmaintenance95388 ай бұрын
Bisa karena ada valve yang agak terbuka atau kurang menutup nya...jadi ada gelembung Udara .... bisa juga bocor di sambungan² selang di chamber nya..bisa juga sumbatan di ruby aperture nya..itu yang paling sering terjadi
@Muslim.25ad Жыл бұрын
Background abnormal with wbc&rbc clog in mindrey 3000 plus.how we remove the error
@d.i.yrepairnmaintenance9538 Жыл бұрын
If rbc and wbc clog..try to check volumetrik board ..cleaning it..if necessary
@d.i.yrepairnmaintenance9538 Жыл бұрын
Look on this video..for volumetrik board..kzbin.info/www/bejne/nnm7oXuQpJJrrqc
@MohamedHassan-pw6xc8 ай бұрын
i have wbcs good but no diff error can u help me plz and if there a way i contact u thanks in advance
@d.i.yrepairnmaintenance95388 ай бұрын
Are all the samples... the diffcount results are not there? try cleaning the ruby stone...and check the exhaust valve from the WBC Chamber
@zaheermaniyar8806 Жыл бұрын
Supper 🎉🎉🎉🎉
@d.i.yrepairnmaintenance9538 Жыл бұрын
Thank you..brother
@akhmadfirzon383 Жыл бұрын
Klo background abnormal terus di plt gimana bang ?
@d.i.yrepairnmaintenance9538 Жыл бұрын
Plt saja..? Nilai nya berapa baground plt nya
@DermawanFaaghnini-mc7pv Жыл бұрын
Hasil wbc 0 trus knapa ya mas?
@d.i.yrepairnmaintenance9538 Жыл бұрын
Hasil pembacaan sample pasient..? Atau baground..?
@DermawanFaaghnini-mc7pv Жыл бұрын
@@d.i.yrepairnmaintenance9538 hasil pembacaan pasien mas
@d.i.yrepairnmaintenance9538 Жыл бұрын
Coba masuk ke service - sytem test/self test..pilih wbc aperture (v) nanti hasil nya berapa voltage nya..kalau terlalu rendah bisa saja board nya..atau wbc bath nya yang rusak
@d.i.yrepairnmaintenance9538 Жыл бұрын
Alatnya bc-3200 juga ya..
@DermawanFaaghnini-mc7pv Жыл бұрын
@@d.i.yrepairnmaintenance9538 oke siap mas
@muzaffarathar1841 Жыл бұрын
Same problem my machine wbc clog back ground abnormal ***********
@d.i.yrepairnmaintenance9538 Жыл бұрын
Try ..zap aperture 3x..after that..flush aperture 3x..and then..pres aspirate without material sample