Thank you so much; very well explain,, also following you on twitter..
@manuelsantander70104 жыл бұрын
Hi Alister! Do you have an extended protocol?
@alistermcneish81482 жыл бұрын
Apologies for the slow reply please contact me via my reading.ac.uk email for a full protocol - it wont be on video.
@北美一家人3 жыл бұрын
You did not do the normalization, could you let me know the reason?
@alistermcneish81483 жыл бұрын
The Normalization is performed using a software extension which is only available in LabChart and associated with the specific equipment we use. We edited out of the video as in the practical class we do not perform that procedure. In reality if the tissue is pretensioned to about 10-15 mN this is the correct range for this tissue
@Alabi.babatunde2 жыл бұрын
What's the difference between normalization and calibration
@alistermcneish81482 жыл бұрын
Calibration is where you set up the equipment to a standardized unit to make sure your read out meets that of a known weight/force (i.e. you set your voltage output to that created by a known force). Normalization is stretching the tissue to a calculated level of force that would be equivalent to 90% of the pressure experience in vivo.