Force FX C Monopolar Calibration

  Рет қаралды 8,293

Better Biomed Channel

Better Biomed Channel

Күн бұрын

Пікірлер: 40
@timw483
@timw483 3 жыл бұрын
Awesome video! I followed along and reccalibrated a problem unit here at my hospital. Thanks so much!
@BetterBiomedChannel
@BetterBiomedChannel 3 жыл бұрын
Thanks Tim
@megabioingenieriambi6498
@megabioingenieriambi6498 3 ай бұрын
Engineer, thank you very much for these videos, they are very helpful. I have a question, I hope you can help me a little more. I did the entire adjustment protocol, the cut was perfect but the coagulation remained the same with a misalignment below the maximum allowed error. What can I do to better adjust the coagulation parameter? Thank you very much.
@drod519
@drod519 3 жыл бұрын
Great video! The QA-ES sure beats using a bank of Dale/Vishay low impedance high power resistors.
@sangeetasang9437
@sangeetasang9437 Жыл бұрын
Thank you and you are the best. Please show some basic of test devices like using Oscilloscope or their features and usage.
@AdeluolaLipede
@AdeluolaLipede 8 ай бұрын
Please send me a video of the COVIDIEN ELECTOCAUTERY MACHINE I CAN USE FOR CHEST SURGERY THANKS
@hoco9665
@hoco9665 9 ай бұрын
Congratulations for the video and the resources you share. I found the adapter in the video interesting but I couldn't find the video in which he presents or creates it. Could you share the link or file for 3D printing, because it looks like it was made that way. A thousand thanks
@mij4604
@mij4604 4 жыл бұрын
Good job. Appreciate
@haelalmesbahi6411
@haelalmesbahi6411 Жыл бұрын
Great vedio. Could you please show us how to calibrate the output of coag . Econ Calibration. Please
@regvalena5972
@regvalena5972 2 жыл бұрын
Good day! I'd like to know how to fix error 212 in force Fx. Thank you so much
@angelcontreras5828
@angelcontreras5828 10 ай бұрын
I have the error 16 and already read the service manual but doesn't have the transformer or inductive resistors. I have the fluke 454A but doesn't have the 30 ohms resistant option, with 100 and 500 would be enough? The error 16 starts when put on the plate on the patient. Also read on a forum that could be "Scaling relays in the control board", how can I probe them? I doesn't have another Force Fx. Some days ago I made the REM calibration because the equipment was showing green REM alarm when plug in the plate even without patient. Also the trouble that made the hospital call me was that the equipment wasn't turning on right, it had a problem in the initial test. When the lights were on bipolar it started as usual but were restarting the autotest when passing trough cut. Should I try to fix It or is better just change the equipment with all that issues? Appreciate any help
@BetterBiomedChannel
@BetterBiomedChannel 10 ай бұрын
Use resistors instead of an analyzer
@rothewall
@rothewall Жыл бұрын
How often does the calibration has to be done to the equipment?
@BetterBiomedChannel
@BetterBiomedChannel Жыл бұрын
Every 6 months we do an output verification. If the results are within tolerance, we won't calibrate anything
@rothewall
@rothewall Жыл бұрын
@@BetterBiomedChannel in this case my hospital do not want to buy analyzers. For this case maybe a calibration per year would be good?
@Guilhermeaugusto-lr5pn
@Guilhermeaugusto-lr5pn 3 жыл бұрын
Good night my friend. Would you have these videos recorded in Portuguese or translated into Portuguese would you have the possibility to send
@haelalmesbahi6411
@haelalmesbahi6411 Жыл бұрын
I have faced error 160 while calibration.
@haelalmesbahi6411
@haelalmesbahi6411 Жыл бұрын
And always fet getting burned during calibration
@josimar4905
@josimar4905 11 ай бұрын
You find a solution?
@haelalmesbahi6411
@haelalmesbahi6411 11 ай бұрын
Yes it was the relay then calibration
@RahulKumar-sz8fq
@RahulKumar-sz8fq 4 жыл бұрын
Thanku so much better biomed
@ricardogutierrezferrer1885
@ricardogutierrezferrer1885 2 жыл бұрын
Excellent!!!!
@mij4604
@mij4604 4 жыл бұрын
From where did you get the adapter plug ?
@BetterBiomedChannel
@BetterBiomedChannel 4 жыл бұрын
I made it. Go watch my video on it from a week ago. I need to make an update video
@haelalmesbahi6411
@haelalmesbahi6411 3 жыл бұрын
I did the calibration for cut and it was perfect . The calibrationb for coag is the seme ??
@BetterBiomedChannel
@BetterBiomedChannel 3 жыл бұрын
Yes.
@haelalmesbahi6411
@haelalmesbahi6411 3 жыл бұрын
And the same values for coag or diffirent ??
@nawarajpaudel6246
@nawarajpaudel6246 3 жыл бұрын
Great video sir. If have not electro surgical analizer (ESA) then possible to calibrate or not esu unit, if can calibrate which tools are required plz help.
@BetterBiomedChannel
@BetterBiomedChannel 3 жыл бұрын
If you have no electrosurgical analyzer it is almost impossible to reliably calibrate or verify function of ESUs..
@nawarajpaudel6246
@nawarajpaudel6246 3 жыл бұрын
Thank you so much for your valuable response.
@r7915
@r7915 4 жыл бұрын
Could you possibly help me out ? I had one of these in today, no rf output on coag - spray or fulgurate but desiccate works perfectly. There is no error codes on boot up & all functions of cut work perfectly. Any ideas & where possibly to look ? & what might have failed ? If not I’ll have to get Medtronic to take a look at it.
@BetterBiomedChannel
@BetterBiomedChannel 4 жыл бұрын
It's no longer supported so the OEM will not assist. Try a different front control panel.
@r7915
@r7915 4 жыл бұрын
This device is still supported by Medtronic in the uk ! I’m just in the process of getting it shipped back for assessment & repair.
@denmachan
@denmachan 2 жыл бұрын
I had a very similar fault and it was the 6 pole relay on the main board. Really easy job to change.
@louislopez2499
@louislopez2499 3 жыл бұрын
sir how can i calibrate if i dont have adaptor plate? can i use the accessories of qaes iii?
@BetterBiomedChannel
@BetterBiomedChannel 3 жыл бұрын
Yes you can calibrate without my adapter. I will have to see my unit at work to deteine the easiest configuration for you.
@kamronrice716
@kamronrice716 3 жыл бұрын
you dont have a QA ES MKII that you could show this process on by chance do you?
@BetterBiomedChannel
@BetterBiomedChannel 3 жыл бұрын
Sorry, I have never used that analyzer
@yashodeepkamble2541
@yashodeepkamble2541 2 жыл бұрын
What if machine showing error 212?
@AQIBRAZA-cp1lf
@AQIBRAZA-cp1lf Жыл бұрын
Control board fault
Personal Tools at Work
7:45
Better Biomed Channel
Рет қаралды 963
Force FX REM Calibration!
9:29
Better Biomed Channel
Рет қаралды 11 М.
Random Emoji Beatbox Challenge #beatbox #tiktok
00:47
BeatboxJCOP
Рет қаралды 48 МЛН
Wait… Maxim, did you just eat 8 BURGERS?!🍔😳| Free Fire Official
00:13
Garena Free Fire Global
Рет қаралды 5 МЛН
ЗНАЛИ? ТОЛЬКО ОАЭ 🤫
00:13
Сам себе сушист
Рет қаралды 3,9 МЛН
Rigel Medical Uni-Therm - ValleyLab Force FX. PM
25:09
Seaward Electronic Ltd.
Рет қаралды 15 М.
Knowledge To Get Ahead In Biomed
18:24
Better Biomed Channel
Рет қаралды 20 М.
QA-ES III Electrosurgical Analyzer Demo Day
31:40
Fluke Biomedical
Рет қаралды 12 М.
Calibrating REM / CQM on Valleylab Force FX C
9:23
Better Biomed Channel
Рет қаралды 6 М.
Syringe Pumps (In a nutshell) with Medfusion 4000
9:06
Better Biomed Channel
Рет қаралды 10 М.
The 90% Toolbag
10:10
Better Biomed Channel
Рет қаралды 4,1 М.
Insufflator Overview
11:42
Better Biomed Channel
Рет қаралды 10 М.
Electro Surgical Unit Overview
34:18
Better Biomed Channel
Рет қаралды 12 М.
The Problem with Biomed 1s
11:50
Better Biomed Channel
Рет қаралды 20 М.
Random Emoji Beatbox Challenge #beatbox #tiktok
00:47
BeatboxJCOP
Рет қаралды 48 МЛН