Awesome video! I followed along and reccalibrated a problem unit here at my hospital. Thanks so much!
@BetterBiomedChannel3 жыл бұрын
Thanks Tim
@megabioingenieriambi64983 ай бұрын
Engineer, thank you very much for these videos, they are very helpful. I have a question, I hope you can help me a little more. I did the entire adjustment protocol, the cut was perfect but the coagulation remained the same with a misalignment below the maximum allowed error. What can I do to better adjust the coagulation parameter? Thank you very much.
@drod5193 жыл бұрын
Great video! The QA-ES sure beats using a bank of Dale/Vishay low impedance high power resistors.
@sangeetasang9437 Жыл бұрын
Thank you and you are the best. Please show some basic of test devices like using Oscilloscope or their features and usage.
@AdeluolaLipede8 ай бұрын
Please send me a video of the COVIDIEN ELECTOCAUTERY MACHINE I CAN USE FOR CHEST SURGERY THANKS
@hoco96659 ай бұрын
Congratulations for the video and the resources you share. I found the adapter in the video interesting but I couldn't find the video in which he presents or creates it. Could you share the link or file for 3D printing, because it looks like it was made that way. A thousand thanks
@mij46044 жыл бұрын
Good job. Appreciate
@haelalmesbahi6411 Жыл бұрын
Great vedio. Could you please show us how to calibrate the output of coag . Econ Calibration. Please
@regvalena59722 жыл бұрын
Good day! I'd like to know how to fix error 212 in force Fx. Thank you so much
@angelcontreras582810 ай бұрын
I have the error 16 and already read the service manual but doesn't have the transformer or inductive resistors. I have the fluke 454A but doesn't have the 30 ohms resistant option, with 100 and 500 would be enough? The error 16 starts when put on the plate on the patient. Also read on a forum that could be "Scaling relays in the control board", how can I probe them? I doesn't have another Force Fx. Some days ago I made the REM calibration because the equipment was showing green REM alarm when plug in the plate even without patient. Also the trouble that made the hospital call me was that the equipment wasn't turning on right, it had a problem in the initial test. When the lights were on bipolar it started as usual but were restarting the autotest when passing trough cut. Should I try to fix It or is better just change the equipment with all that issues? Appreciate any help
@BetterBiomedChannel10 ай бұрын
Use resistors instead of an analyzer
@rothewall Жыл бұрын
How often does the calibration has to be done to the equipment?
@BetterBiomedChannel Жыл бұрын
Every 6 months we do an output verification. If the results are within tolerance, we won't calibrate anything
@rothewall Жыл бұрын
@@BetterBiomedChannel in this case my hospital do not want to buy analyzers. For this case maybe a calibration per year would be good?
@Guilhermeaugusto-lr5pn3 жыл бұрын
Good night my friend. Would you have these videos recorded in Portuguese or translated into Portuguese would you have the possibility to send
@haelalmesbahi6411 Жыл бұрын
I have faced error 160 while calibration.
@haelalmesbahi6411 Жыл бұрын
And always fet getting burned during calibration
@josimar490511 ай бұрын
You find a solution?
@haelalmesbahi641111 ай бұрын
Yes it was the relay then calibration
@RahulKumar-sz8fq4 жыл бұрын
Thanku so much better biomed
@ricardogutierrezferrer18852 жыл бұрын
Excellent!!!!
@mij46044 жыл бұрын
From where did you get the adapter plug ?
@BetterBiomedChannel4 жыл бұрын
I made it. Go watch my video on it from a week ago. I need to make an update video
@haelalmesbahi64113 жыл бұрын
I did the calibration for cut and it was perfect . The calibrationb for coag is the seme ??
@BetterBiomedChannel3 жыл бұрын
Yes.
@haelalmesbahi64113 жыл бұрын
And the same values for coag or diffirent ??
@nawarajpaudel62463 жыл бұрын
Great video sir. If have not electro surgical analizer (ESA) then possible to calibrate or not esu unit, if can calibrate which tools are required plz help.
@BetterBiomedChannel3 жыл бұрын
If you have no electrosurgical analyzer it is almost impossible to reliably calibrate or verify function of ESUs..
@nawarajpaudel62463 жыл бұрын
Thank you so much for your valuable response.
@r79154 жыл бұрын
Could you possibly help me out ? I had one of these in today, no rf output on coag - spray or fulgurate but desiccate works perfectly. There is no error codes on boot up & all functions of cut work perfectly. Any ideas & where possibly to look ? & what might have failed ? If not I’ll have to get Medtronic to take a look at it.
@BetterBiomedChannel4 жыл бұрын
It's no longer supported so the OEM will not assist. Try a different front control panel.
@r79154 жыл бұрын
This device is still supported by Medtronic in the uk ! I’m just in the process of getting it shipped back for assessment & repair.
@denmachan2 жыл бұрын
I had a very similar fault and it was the 6 pole relay on the main board. Really easy job to change.
@louislopez24993 жыл бұрын
sir how can i calibrate if i dont have adaptor plate? can i use the accessories of qaes iii?
@BetterBiomedChannel3 жыл бұрын
Yes you can calibrate without my adapter. I will have to see my unit at work to deteine the easiest configuration for you.
@kamronrice7163 жыл бұрын
you dont have a QA ES MKII that you could show this process on by chance do you?