hubs.ly/H07K8Sp0 The Hanna Instruments pH, Conductivity, TDS Low Range Combo Tester is a simple and reliable tool for anyone who needs to take multiple measurements in a variety of samples.
Пікірлер: 65
@Arycke6 жыл бұрын
I used this in field research and it worked wonderfully. The tolerance is nominal and it is quite portable. Thank you for the upload and product.
@rundowner462 жыл бұрын
Good morning kassandra. I just bought model #HI98129 PH/EC/TDS combo tester. Reading the manual I couldn't figure out how to change, the temperature units of measurement. This presentation video, made it very easy to understand. I kept press but not, holding mode button. For 6 seconds after the unit was turned off. Thank you for the presentation video. Now I need some distilled water.
@ChrisGtek2 жыл бұрын
I’ve had my Hanna combo Meter since 2011 haven’t looked back since i even own a blue lab combo one I prefer my Hanna meter anytime 👍👍👍
@peaches0212724 жыл бұрын
Simple to follow with this video.
@haidermehdi13224 жыл бұрын
good evening can you tell me how to calibrate the hanna H199300 consuctivity meter
@agassociates84864 жыл бұрын
Just bought your HI98129. I'm unable to calibrate it even after following all the steps. ph 7.01 solution shows 7.18 and ph 4.01 solution shows 4.57. Can you guide?
@jagdishraval74663 жыл бұрын
My new instrument is autometic goes in 10.01 Ph Calibration even on selecting 7.01 as well using 7.01 buffar
@Babykuttyvibes4 жыл бұрын
Trying to do the two point calibration with brand new HI98129. Always getting the WRNG message when i am switching from 7.01 buffer(HI7007) to 4.01buffer(HI7004). I rinse the electrode in RO water before putting it in 4.01 buffer. What must be going wrong? Please advice..
@moritzkeller45023 жыл бұрын
I am wondering as well, nothing about this errorcode in the manual. Have you found a solution?
@clutch11263 жыл бұрын
Still no answer... Crap giving me the same thing
@redrhapsody1415 Жыл бұрын
Try leaving the probe in your solution for a bit longer. The probe says wrng when it doesn't detect the correct solution. After a few seconds the error message should go away allowing you to proceed. Hopefully this helps you in your endeavors!
@malte49195 жыл бұрын
Really like my Hanna combo, but can't your company give names which I can remember? What do I need to calibrate? 1805226 or 2590522 or was it 598643? :(
@HannaInc5 жыл бұрын
Hi Bee. Glad you like your meter! We do have some long names for our products, but hopefully we can help simplify it for you. If you visit hannainst.com you can find all the calibration solutions you need. Typically with this meter, you would want pH 7.01 Buffer (HI7007) pH 4.01 buffer (HI7004) pH 10.01 Buffer (HI7010) & 1413 µS/cm EC Standard (HI7031) Depending on what you're measuring, you may not need the pH 10.01 buffer. It all depends on your need. Hope this helps!
@NatureBound4 жыл бұрын
I'm on my third one this year. My last one lasted two weeks, never dropped, never went dry, used and cleaned daily. kinda getting sick of spending $200 at a time for something that can't last more than a few months.
@kbg4life074 жыл бұрын
Dang mine has been running strong for 7 years already
@fernandoolalde62313 жыл бұрын
@@kbg4life07 😐
@wayneessar74892 жыл бұрын
Isn't there a two year warranty?
@heywoodjablome5630 Жыл бұрын
@@wayneessar7489 just one year, but no warranty against careless or clumsy users.
@jojimonty25066 жыл бұрын
Where can I get the test solutions? My meter is not reading correctly.
@HannaInc6 жыл бұрын
Hi Joji. You can visit a local retailer or visit hanna's website for buffers and cleaning/storage solutions. How old is your meter? Can you give me a little bit more information about what's going on?
@clutch11263 жыл бұрын
Bought one of these but when calibrating it tells me wrng when put on to 4.01 solution. Cleaned for over an hour and soaked in 7.01...
@redrhapsody1415 Жыл бұрын
Hopefully you found a solution! What I did when this problem occurred was rinsing with DI water in between cal solution. It will still say wrng when you go to place the probe into the next solution but, ignore that and leave it in the next solution for a minute and it should continue with calibrations. The probe will say wrng when it doesn't detect the correct solution and after a few seconds it should allow you to proceed. Hopefully this helps the next person who has this problem!
@nawelder7686 жыл бұрын
PLEASE I WOULD LIKE TO KNOW HOW TO CHANGE THE POINT OF CALIBRATION OF CONDUCTIVITY BECAUSE 1413 IS TO HIGH
@HannaInc6 жыл бұрын
Hi nanou. You want to be in EC/TDS mode. From there you can enter calibration mode and change the setting from there. Hope this helps! You can find the demonstration at 3:47 in the video.
@myneymo773 жыл бұрын
My left button, only switch off the unit, anyone can help? No mode button functioning
@rundowner462 жыл бұрын
I had the same problem buddy. Until I watched the video. I'm a slow one. Took me a few to many off a one. To understand; how to get there. I turned mine on. Then made sure it was in ph mode. Pressed the mode button. Which will show off. Now hold for 6 seconds. Changed the temperature range. The instructions in the manual were confused. But kassandra cleared it up for me. Thank you kassandra. I hope your doing well. Since all this covid $h!t. Hope this helps Le Badger? I hope all is well buddy.
@miguelcarrera835 жыл бұрын
Hello Can you help me? I can't calibrate my HI98129. I'm using a new 1413 microsiemens solution, but the screen says "Wrong" My Hi98129 is new too. What could be the problem?
@HannaInc5 жыл бұрын
Hi Miguel. It sounds like you're in the pH mode. The first thing you want to do is put the meter in EC mode, and then enter Calibration.
@miguelcarrera835 жыл бұрын
@@HannaInc No, I'm in the EC mode. Another thing? Thank you.
@HannaInc5 жыл бұрын
@@miguelcarrera83 Are you in µS or ppm mode. Is it asking you for the 1413 or 1382 when you enter CAL mode?
@miguelcarrera835 жыл бұрын
@@HannaInc In microsiemens mode. I have 0.5 of conversion factor and 2.0% of temperature coeficient
@HannaInc5 жыл бұрын
@@miguelcarrera83 contact tech@hannainst.com They can help you step by step to see what's happening. Hope this helps!
@matutawavitabernard17026 жыл бұрын
need how to i can make calibration
@HannaInc6 жыл бұрын
Hi MATUTA, calibration is really simple. Press the MODE button until you see CAL on the screen. From there, it will prompt you for your buffers.
@serpico22064 жыл бұрын
Why dont you guys deliver “calibrated” version of this product ? its 150 $.... cant you guys do “pre calibrated” electrodes?
@dennfox14644 жыл бұрын
Best question on this page!
@tee2282 жыл бұрын
i think its calibrated out the factory but after being stored on store shelves for who knows how long, the probe could possibly dry out a bit making it not as accurate
@raymondolivares10935 жыл бұрын
98129 test TDS as well?
@HannaInc5 жыл бұрын
Hi Raymond. Yes it does!
@naveedakhtarid2 жыл бұрын
how to remove wrng error during ph calibration process.
@naveedakhtarid2 жыл бұрын
The answer of this question is still pending..... Plz any expert in calibration help
@redrhapsody1415 Жыл бұрын
The meter gives the error message "wrng" when it doesn't detect the correct solution. After a few moments it should go away allowing you to proceed with your calibrations. This occurred for me when I would rinse my meter with DI water after the 7.01 solution prior to putting it into the 4.01pH solution. Hopefully this helps!
@solangeriveragarces3136 жыл бұрын
you have an email for information?
@HannaInc6 жыл бұрын
You can contact sales@hannainst.com or visit hannainst.com for more info!
@harmeetsingh3715 жыл бұрын
Hello maam one of my ec/ph meter didnot calibrate model hanna HI98130
@HannaInc5 жыл бұрын
Hi Harmeet. Would it not enter calibration mode or not stabilize? Many times, a stabilization issue is due to a dehydrated electrode. It's important to keep it stored in storage solution. Something that works is if you let your electrode soak in storage solution for an hour, it will rehydrate and should work just fine.
@g_pokesport91656 жыл бұрын
My meter looks exactly like this but reads in ph, ms, ppt.
@HannaInc6 жыл бұрын
Hi Gary! You must have the HI98130 meter. Awesome choice!
@jakeyasutomi6 жыл бұрын
Holding the mode button only turns my meter on. Is there anything else I can do to calibrate it ?
@HannaInc6 жыл бұрын
Hi Jake, The Mode button does turn the meter on, but once it's on, press the button again and hold it. You'll see CAL displayed on the screen. Release the button and it will ask you for your first calibration buffer. If you still have issues let me know!
@emilianobernardo42856 жыл бұрын
I just want to ask about the calibration of the High range ph meter I was confused while I'm doing calibrate using buffer ph 7.01 then after a few Seconds 10.01 was indicated in LCD screen instead of 4.01. it's there's any problem ? or I can use also the 10.01 for the 2nd point of calibrating?.
@HannaInc6 жыл бұрын
The tester will automatically recognize which buffer you use. It seems it is just going to the next point up, but if you were to use pH 4.01, it would recognize and calibrate accordingly. Using pH 4.01 or 10.01 first will do a 1 point calibration by default.
@gabesworld78302 жыл бұрын
paid like 150 for this is I remember correctly and after less than a year it will not calibrate. Sucks!
@naveedakhtarid2 жыл бұрын
Plz,
@kennethtomer55535 жыл бұрын
Stay away from these. I bought one for christmas and did not work right out of package. And the warranty is not covered because hydro store did not give me a receipt.. what a waste
@HannaInc5 жыл бұрын
Hi Kenneth. Buying from a store can sometimes mean the meter may have been sitting for a while. Electrodes need to be rehydrated with storage solution when they become dry. I'd recommend soaking it in your storage solution for 2 hours and then trying it again. If it still doesn't work, contact tech@hannainst.com
@kennethtomer55535 жыл бұрын
Not when it won’t even power on
@DontWorryImNorwegian5 жыл бұрын
You don't show/tell people how to get the damn thing into calibration mode before showing how to calibrate it!
@HannaInc5 жыл бұрын
Hey there! Not to worry. If you press the mode button for 3 seconds it will bring you into calibration mode.