@doctor6932
@doctor6932 6 ай бұрын
The video fails to mention image intensifiers are lead lined inside the aluminum housing and the aluminum input faceplate (about 1mm thick) that separates the vacuum envelope is where the x-ray enters. Also it should be pointed out the grid is totally separate from the image intensifier and not sold with it because each application requires its own parameters of a grid.
@HowRadiologyWorks
@HowRadiologyWorks 6 ай бұрын
Thanks Dr for the good points. We do mention the outer housing at 3:14 but not to that detail and it is a good point that grids are separate and we have other videos on grids as well
@emilyzeno7199
@emilyzeno7199 Жыл бұрын
I have a question for you... if have two generators from different brands and they both have the same mA station what might cause one to be slightly more powerful than the other? Could there be more keV in one machine than the other? Thanks
@HowRadiologyWorks
@HowRadiologyWorks Жыл бұрын
Different generators could be generating different kVp waveforms and the x-ray spectrum also depends on other factors like the anode material and pre-patient filtration.
@goodluckspeshi7868
@goodluckspeshi7868 Жыл бұрын
Very helpful, thank you ♥️ 🇳🇬
@UraniumGirl
@UraniumGirl Жыл бұрын
Thanks, it was very helpful.
@brianmaunda5493
@brianmaunda5493 4 ай бұрын
This was helpful
@HowRadiologyWorks
@HowRadiologyWorks 4 ай бұрын
Thanks Brian, that’s our goal. Thanks for leaving a comment
@medicalengcenter6215
@medicalengcenter6215 Жыл бұрын
And how can I make sure of high tension power of ii tube
@doctor6932
@doctor6932 6 ай бұрын
In practicality you cant. The high voltage requires HV probes. The reality is you measure test points on the image intensifier power supply which are scaled down versions of the actual power being applied. Remember this is a vacuum tube and there isn't any loading effect on the power supply.
@qzorn4440
@qzorn4440 Жыл бұрын
A very nice overview of an image intensifier. I would like to replace a 30 year old 12" Image Intensifier coupled to a CCD camera. Which brand of equipment are you using? 😎 Thank you.
@HowRadiologyWorks
@HowRadiologyWorks Жыл бұрын
Sorry I don’t have brand recommendations
@qzorn4440
@qzorn4440 Жыл бұрын
@@HowRadiologyWorks Thank you.
@doctor6932
@doctor6932 6 ай бұрын
Thales is the most common brand but Dunlee is cheaper and still high quality. Phillips is high quality. Just about any image intensifier is good. Key factors are gain and resolution, contrast ratio, etc.
@medicalengcenter6215
@medicalengcenter6215 Жыл бұрын
What black image mean in the monitors?
@HowRadiologyWorks
@HowRadiologyWorks Жыл бұрын
Is the whole image black or do you mean the black parts of the image
@doctor6932
@doctor6932 6 ай бұрын
You must first separate the Image intensifier output from the camera. First of all verify you have dose (i.e. 30uGy for 16", 50uGy for 9", 80uGy for 6", etc. Inverse dose to size due to minification.). Then verify light output from image intensifier. It might appear dim but don't worry about that. If you do then it is in your camera system. Most commonly the problem is the image intensifier power supply but the image intensifier cant be ruled out.
@medicalengcenter6215
@medicalengcenter6215 Жыл бұрын
You really great prof
@HowRadiologyWorks
@HowRadiologyWorks Жыл бұрын
Thanks for the props
@ramjoshi4019
@ramjoshi4019 7 ай бұрын
can u explain about veiling glare plz..
@HowRadiologyWorks
@HowRadiologyWorks 7 ай бұрын
Okay I’ll put it on the list
@doctor6932
@doctor6932 6 ай бұрын
Veiling glare is reflected light that occurs from reflections of the light at different interfaces. Basically light reflections inside the output port and between the camera and output port. For instance an output port glass window (or one with a poor transmission factor) would cause more veiling glare than a thin output window. Because the light would go into one side of the glass through the glass and instead of it all exiting the other side some would be reflected backward at an angle to the input side again and then get reflected again. That is the primary cause. This causes a lower gain and lower contrast ratio.
@abdovitamins6331
@abdovitamins6331 Жыл бұрын
Good work.👌 Greetings
@HowRadiologyWorks
@HowRadiologyWorks Жыл бұрын
Thanks Abdo vitamins.
Fluoroscopy
25:10
Rock The Registry
Рет қаралды 31 М.
Fluoro Physics Goodenberger
32:45
Neil Hansen
Рет қаралды 4,9 М.
The Giant sleep in the town 👹🛏️🏡
00:24
Construction Site
Рет қаралды 20 МЛН
Happy birthday to you by Tsuriki Show
00:12
Tsuriki Show
Рет қаралды 12 МЛН
А ВЫ УМЕЕТЕ ПЛАВАТЬ?? #shorts
00:21
Паша Осадчий
Рет қаралды 1,5 МЛН
Fluoroscopy Magnification and Pulsed Fluoroscopy
13:02
How Radiology Works
Рет қаралды 3,8 М.
How an X-Ray Tube Works⚡How X-Rays are Generated
9:45
VirtualBrain [ENG]
Рет қаралды 18 М.
RADT 086 Image Intensification Tube
12:39
christyfoster2002
Рет қаралды 37 М.
The Image Intensifier Tube
5:29
Jeremy Enfinger
Рет қаралды 33 М.
Tomosynthesis (Does it offer 3D Mammography)
15:02
How Radiology Works
Рет қаралды 6 М.
Fluoroscopy
5:40
Radiologic Sciences
Рет қаралды 19 М.
Image Intensifier tips!
6:02
Meaghan Piretti
Рет қаралды 9 М.
History of X-rays
24:21
Doctor Klioze
Рет қаралды 273 М.
Radiation Measurements Overview - X ray production and Safety
6:19
Clover Learning
Рет қаралды 12 М.
The Giant sleep in the town 👹🛏️🏡
00:24
Construction Site
Рет қаралды 20 МЛН