Thank you Better BioMed. You're doing a great job educating people with right knowledge. :)
@rn87mom943 ай бұрын
Using the AVI plums since 1988. Before then, we had iVacs. They had this laser that you’d attach onto the drip chamber that counted the drops.
@bernardlum44678 ай бұрын
Please do a video on CADD Solis PCA pump by Smith Medical
@daltondelaney95342 жыл бұрын
You did make one video where you broke down an Alaris 8100 along with a little patient monitor and a vacuum regulator. I remember watching it before I started helping out with one of the hospitals in my district for IV pump month it was super helpful because you tore it down completely and we’re really thorough.
@sangeetasang9437 Жыл бұрын
Agreed. You're very cleat and precise. You know the PULSE and what's needed...
@orlandodiaz40592 жыл бұрын
Very good 👍 nice information
@josecampo52072 жыл бұрын
Buén video amigo. Muchas gracias. ¿Sabes como ajustar la bomba de infusión Medcaptain MP 60? Es que me está bombeando más de lo programado. Saludos.
@sonalsandeepchand16082 жыл бұрын
Hi, thank you for sharing the informative video. Do you have any repair videos on Hospira plum A infusion pumps?? We have a few at our facility and we end up sending it to Hospira service center for repairs..
@BetterBiomedChannel2 жыл бұрын
I don't yet but now I will.
@sonalsandeepchand16082 жыл бұрын
@@BetterBiomedChannel thanks and much appreciated 🙏
@scarops2 жыл бұрын
This was helpful. Thank you.
@jp3d2k Жыл бұрын
I believe Alaris is a linear peristaltic pump.
@yankoalejandromartinezamay79662 жыл бұрын
Hello, I have a problem with an infusionplum A+ pump, brand "N184 Prox Occl A/ Air. Do you know how to repair it? The line is not clogged.
@BetterBiomedChannel2 жыл бұрын
There are 2 plastic lobes on the disposable cassette which are used to detect air in line. There is a clear / white tiny button insider the cassette tray which is pressed to clear and reset any errors.
@yankoalejandromartinezamay79662 жыл бұрын
@@BetterBiomedChannel I already press the button and the equipment erases the errors, but when I put the cassette back in, the same error "N184 Prox Occl A/ Air" comes out. the pump has 2 plastics inside the cassette, one white and one black. black is broken. I think it's the occlusion sensors. I have a photo, I can send it to you by mail please help.
@salembajrami90192 жыл бұрын
Can you do pm and cm videos on this?
@BetterBiomedChannel2 жыл бұрын
Sorry, I won't do PM videos. Too much liability
@salembajrami90192 жыл бұрын
@@BetterBiomedChannel Just started working as a biomed 1 and sometimes the manuals are overly complicated and it is really hard to find good pm videos. But anyways appraciate a lot what you do!
@350zLeMans2 жыл бұрын
Hi any recommendations on where I can find some good study material for BMET1?
I am trying to study for the cbet. Any idea on the best source of study material?
@BetterBiomedChannel2 жыл бұрын
There are several study guides available but cbet study groups might help more. I personally like the flash cards to find your strengths and weaknesses
@derrick53802 жыл бұрын
@@BetterBiomedChannel thanks, so I guess any of them will do. I was googling stuff and there were so many different ones. I did see a site with a book and flash cards though so thanks