I love how this guy presents the information :) makes me feel joyful
@kpmitton Жыл бұрын
NOTE TO VIEWERS the data plotted must be Ct values versus the LOG of the DNA concentration (or log of the DNA dilution relative to the concentrated standard)
@dalmironbonnin10 жыл бұрын
all your vids are very helpful! thanks!
@catherineokoi17314 жыл бұрын
You are a pro! Thanks for the explanation
@bernhardandresen7 жыл бұрын
Why didn't I find you Videos a year ago!? :(( I would have needed them so bad.
@kpmitton Жыл бұрын
When you dilute a representative DNA sample to make a dilution series for checking the efficiency of a PCR reaction, include dilutions that will include the dilutions you will also use of your actual samples. Note that if you use the slope to calculate Efficiency using just the first more concentrated samples you may get a value that is over 100%. Like 110%. But you cannot actually have more than perfect doubling 100% efficiency. So why do you get >100%? Simple, that is because you have inhibition of PCR in the most concentrated sample and as you dilute the sample the inhibiting factors decrease, which causes the efficiency to increase from its lower, inhibited, efficiency. This gives you a slope and calculation of >100%. You can check the slope using just the more diluted standards and will find the PCR efficiency does not exceed 100%. If possible, dilute your test samples to use the range of dilutions that avoid the inhibited PCR efficiency. It will just take two or three dilutions to often remove this problem. Also, no matter what your dilution is, find out the PCR efficiency around that dilution region and consider using the full equation that includes adding the fractional PCR efficiency for both your gene of interest and normalizer gene PCRs. You can still use the reactions for quantitative analysis even if they are lower efficiency, like 75%, as long as you use calculations that include the efficiency. The simple Delta-Delta Ct method equation is just the simplified form of the equation when the fractional PCR efficiencies are 1 (100%). Ken Mitton, PhD FARVO. Eye Research Institute, Oakland University, Michigan.
@harkamanghag68779 жыл бұрын
Great Video, got to discuss PCR with Breast Cancer Surivor's at San Francisco Avon Walk at Fort Mason. They Love the fact the technology could possibly lead to a cure.
@emc17602 жыл бұрын
Is it possible to import a standard curve from a previous PCR run on the Ab7500? and use it in another assay?
@yenesewmihret98992 жыл бұрын
Dear Taqman, I run 3 kDNA PCR with its standard series, when i did the first run the standard was ok, meaning the difference b/n two Ct values was 3.2..... but the run's of standard were inconsistent. so my question was can I calculate the unknown sample by comparing it along with the first st curve. Thank you!!!!
@nikhilponnam683010 жыл бұрын
what if my slope is -4.4 should i have to repeat the std curve analysis by making new dilutions or is their any other way to minimize the error in std curve
@大错特错不要来侮辱我7 жыл бұрын
Hi. May I measure the PCR efficiencies of all pairs of primers in one PCR plate? Separately from my unkonw samples.
@wondetsegaye40434 жыл бұрын
what is the difference between dilution series and serial dilution? could you explain how to quantify DNA for forensic human identification? thank you
@thermofisher4 жыл бұрын
Thank you for your question. A serial dilution is when you start with tube 1 and dilute it into tube 2. You then take tube 2 and dilute it into tube 3 and so forth. The terms dilution series and serial dilution can often be used interchangeable. If you have questions about your experimental setup, please contact us at thermofisher.com/askaquestion. Thank you.
@gertrudehemington757010 жыл бұрын
What is an appropriate method for determining and omitting outliers?
@gertrudehemington757010 жыл бұрын
Super, thanks!
@贾潇潇5 жыл бұрын
OMG thx so much for this hint!
@Kadishmall11 жыл бұрын
What sofware is that at minute 1:09? thanks
@lifetechnologiescorp854411 жыл бұрын
That is the StepOnePlus software that comes with the instrument. Hope that helps.
@jdevolanyc10 жыл бұрын
What is considered "an acceptable standard material?"
@shajidislam22228 жыл бұрын
can you please explain the slope value..how its( -3.32)
@thermofisher8 жыл бұрын
The slope comes from the equation y = mx + b, which is plotted through the points generated by the efficiency check curve. The points on the x-axis are log [input cDNA], and the y-axis are the Ct values.